4.
Press
std
to access the Standardize mode.
The selected buffer group is displayed briefly.
2 4 7 10 12
BUFFER
select
5.
Do wait for reading to stabilize.
6.
Press
std
again to initiate standardization. The meter will
automatically recognize the buffer.
pH
Standardize
and then meter returns to the Measure screen.
)
Press
mode
key while in the standardization mode to return to
measurement without accepting calibration.
31