Fisher Scientific 88861023 Operation Manual Download Page 8

The front panel of the Microplate Shaker contains all the controls needed to operate the unit.

TIME

This chapter covers the control panel and its operation.

1. Start/Stop button: Start or stop the instrument.
2. RUN indicator light: The light is on when the instrument is
running and off when the instrument is in standby.
3. TIME display window: The window shows cumulative time
(in continuous mode) or remaining time (in timer mode). The
range of time displayed is 0 to 99 hours and 59 minutes. The
accuracy is 1 minute. 
4. SPEED display window: The window shows set speed
(when the instrument is in standby) or current speed (when the
instrument is running). 
5. Set Time Buttons: UP/DOWN Arrow buttons are used to
increase/decrease the set time of the instrument.

. Set Speed Buttons: UP/DOWN Arrow buttons are used to

increase/ decrease the set speed of the instrument.

6

1. Connect all the components according to the figures shown
on page 4 of this manual. Use grounded power outlet.
2. Press down the power switch on the back right side of the
instrument and put it to the “|” state and then the instrument is
in standby.

Section 3 Operation 

3.1 Control Panel

3.2 Installation

3.3 Settings

Time Settings

1. Continuous mode

Press the “     ” or “     ” arrow button below the TIME display
window. When the number shown on the display window starts
flashing, press “     ” arrow button to decrease the time to 00:00
and then release the button. The time setting is finished after the
number shown on the display window has flashed twice.

2. Timer mode

window. When the number shown on the display window starts
flashing, press “     ” or “     ” arrow button to increase or
decrease the time value. Release the button when the time
shown on the display window reaches the set value. The time
setting is finished after the number shown on the display window
has flashed twice.

Speed Settings

Press the “     ” or “     ” arrow button below the SPEED display
window. When the number shown on the display window starts
flashing, press “     ” or “     ” arrow button to increase or
decrease the speed value. Release the button when the speed
shown on the display window reaches the set value. The speed
setting is finished after the number shown on the display window
has flashed twice 

Note: 

press the “     ” or “     ” arrow button for a longer time to

accelerate the setting.

Run and Stop

1. Continuous Mode

Press

button and the instrument will start running with the

specified settings and the RUN indicator light will be on. The
TIME display window will show the cumulative time and the
SPEED display window will show the current speed. Press “     ”
button again and the instrument will slow down until it stops. 
The instrument will then be in standby and the two display
windows will show the set values.

“     ”

2. Timer Mode

Press 

 button and the instrumentwill

The

TIME display window will show the remaining time and the
SPEED display window will show the current speed. Press 
button again and the instrument will slow down until it stops.
The instrument will then be in standby and the two display
windows will show the set values.

“     ”

“     ”

 start running with the

specified settings and the RUN indicator light will be on. 

7

Press the “     ” or “     ” arrow button below the TIME display

Summary of Contents for 88861023

Page 1: ...Microplate Shaker Operations Manual V1 12 21 17 ...

Page 2: ...This manual covers the model shown below 1 NA Model Voltage Description 88861023 88861024 100 240V 100 240V Microplate Shaker US plug Microplate Shaker Intl plug EU Model N A 15504070 ...

Page 3: ...or transmitted in any form or by any means including photocopying recording or using information storage and retrieval systems for any purpose other than the purchaser s own use without the express written permission of the manufacturer Any other product names and services identified in this manual are trademarks or registered trademarks of their respective owners No such use or the use of any tra...

Page 4: ...tions Environmental Conditions Safety Instructions Capacity and Speed Section 3 Operation Control Panel Settings Installation of Accessories Section 4 Safety Tips and Maintenance Section 5 Troubleshooting Section 6 Optional Accessories Spare Parts Section 7 Warranty 2 Installation 4 4 4 4 5 5 5 5 6 7 7 7 7 8 9 10 11 12 ...

Page 5: ...iver Has been installed on the instrument 1 1 Packing List Table 1 Packing List 6 12 N A N A N A Microplate Shaker 1 1 1 M3 6 Screw 12 6 N A 1 1 1 1 1 Figure 1 2 Connections Main Body Power Cable all the components before discarding the packing If there are any missing or damaged pieces please contact the manufacturer immediately 3 Place the instrument on a level and firm surface to avoid vibratio...

Page 6: ... at an appropriate speed to avoid spillage of liquids Note These max speeds are just for reference We did the tests on the marble surface Its friction coefficient is 0 32 1000rpm 800rpm Load Type 96 well Microplate 96 well Microplate Liquid Capacity 1 2 well deep 2 3 well deep Max Speed Fixed Tube Rack 40 Φ16 Fixed Tube Rack 40 Φ14 Adjustable Tube Rack 40 Φ14 Adjustable Tube Rack 40 Φ16 500rpm 500...

Page 7: ...rdous atmosphere or with hazardous materials for which the unit was not designed Also the user should be aware that the protection provided by the equipment may be impaired if accessories used are no provided or recommended by the manufacturer or are used in a manner not specified by the manufacturer CAUTION To avoid electrical shock completely cut off power to the unit by disconnecting the power ...

Page 8: ...ow has flashed twice 2 Timer mode window When the number shown on the display window starts flashing press or arrow button to increase or decrease the time value Release the button when the time shown on the display window reaches the set value The time setting is finished after the number shown on the display window has flashed twice Speed Settings Press the or arrow button below the SPEED displa...

Page 9: ...the machine will stop running press any key to put the instrument in standby 3 After the timer goes off the buzzer will alarm and the instrument will stop running automatically Note To ensure shaking operation smooth and steady it may take 1minute for the microprocessor control system to accelerate the tray to the set speed Finish Operation After the operation is finished press the power switch at...

Page 10: ... motor It is maintenance free and has a long service time high quality and low noise level b Surface can be cleaned with a mild detergent and water If accidental spillage of liquids caused by mishandling or contained breakage occurs on the surface of the instrument please shut down the instrument and clean up the liquid immediately If the liquid has already spilled into the unit cut off the power ...

Page 11: ...unction Tray loose Microplate loose Please refer to the following table to troubleshoot if any malfunction occurs If the problem still exists contact your local sales representative Keep record for maintenance Solution Connect the power Turn power switch on Replace power adaptor Adjust the weight and position of load decrease rotation speed Contact Fisher Scientific Contact Fisher Scientific Faste...

Page 12: ... Plug Power Cable CN Spare Part Plug Power Cable EU Spare Part Plug Power Cable UK Spare Part Plug 1 1 1 1 1 15554120 15564120 15574120 Fixed Tube Rack 40 Φ14 Fixed Tube Rack 40 Φ18 Fixed Tube Rack 27 Φ22 Fixed Tube Rack 21 Φ30 88861148 N A N A N A 88861149 88861150 88861151 N A N A N A 262 112 140mm 262 112 140mm 262 112 140mm 262 112 140mm 2 2 2 2 2 2 2 262 112 140mm 262 112 171mm 262 112 171mm ...

Page 13: ... and maintenance this product is warranted for 24 months against defective materials or workmanship The 24 month warranty period begins from the delivery date of this product For product quality or performance issues contact Fisher Scientific Customer Service Section 7 Warranty 12 ...

Page 14: ...9 478 it fishersci com Finland 358 0 9 8027 6280 fishersci fi France 33 0 388 67 14 14 fishersci fr North America Canada 1 800 234 7437 fishersci ca Netherlands 31 0 20 4887 70 00 nl fishersci com Norway 47 22 95 59 59 fishersci no Portugal 351 21 425 33 50 pt fishersci com Spain 34 002 239 303 es fishersci com Sweden 46 31 68 94 30 fishersci se Switzerland 41 0 56 618 41 11 ch fishersci com UK 44...

Reviews: