Selecting a Location
Choose a location for the refrigerator/freezer that will provide
a clearance of 4 in. at the top, 8 in. at the rear and 3 in. at
each side.
Appropriate electrical power must be available. Locate the
refrigerator/freezer within eight feet of the power outlet so that
no extension cord is required. Attach the refrigerator/freezer to
the facility’s electrical supply as directed by the National
Electrical Code Article 440
.
Leveling the Unit
This refrigerator/freezer must be level in order to provide
adequate condensate drainage as well as proper door
alignment and operation. The refrigerator
/freezer
should be
in its final operating location and set so that it is firmly
positioned on the floor.
12
Wiring
General purpose units use 15 or 20 amp power cords. They
are rated for 115 volts, AC, 60 Hz. Do not use extension
cords, and always use a three prong grounded wall outlet.
Figure
3
shows standard NE service cord plugs and wall
outlets used for
ou
r
refrigerators and freezers.
Figure 3. NEMA Plugs and Receptacles
Shelves
For 20ER:
Shipped inside each cabinet are four shelves, one shelf is a
sheet metal shelf & others are wired shelves that are fixed.
For 20EF:
Shipped inside each cabinet are four fixed wire shelves
which also serve as part of evaporator coils
.
I
NSTALLATION