Alarm Relays (Not used on this model) 

The SPST 24V 500mA relay contacts are normally open and will close 
on alarm conditions when the alarms are enabled.  Relay contacts are 
always functional when the alarm is enabled.  Dip Switch #8 has no 
effect on the relay.  The relay will close only during minimum and 
maximum alarm conditions. 

Display Symbols 

C

 - Indicates 

°

F

 - Indicates 

°

O

 - Zero calibration mode. 

A

 - Alarms enabled or alarm set mode. 

MIN

 - This symbol is displayed when you are setting a minimum alarm 

or during a minimum alarm condition when alarms are enabled. 

MAX 

- This symbol is displayed when you are setting a maximum alarm 

or during a maximum alarm condition when alarms are enabled. 

LED

 - The LED will be green whenever the unit is activated.  The LED 

will flash green during initialization and warm-up.  The LED showing red 
indicates the unit is in calibration mode.  The LED turns orange during 
shut down. 

Dip Switch Set-Up 

To set-up the recorder for your specific application, you might need to 
change some of the Dip Switches.  The Dip Switches are located 
underneath the chart.  Use a pen or small screwdriver to flip the 
switches.  Switches 1 and 2 not used, set to off. 
 Factory switch settings 1-8:  00011010  (0 = Off, 1 = On) 

Recording Time 

The recorder has two different recording options: 24 hour and 7 day.  
Dip switch #3 controls the recording time. 

1 Day  

 #3 On 

7 Day  

 #3 Off 

°

F/

°

You can record in 

°

F or 

°

C with by changing Dip Switch #4 

°

 #4 On 

°

 #4 Off 

Temperature Range 

The temperature range is controlled by Dip Switches #5, #6 & #7. 

-50 to +50 

 #5 On 

  

#6 

Off 

 #7 

On 

0 to 50 

 #5 On 

  

#6 

Off 

 #7 

Off 

0 to 100 

 #5 Off 

  

#6 

Off 

 #7 

Off 

 
Alarm Speaker (Not used on this model) 

The alarm speaker may be activated or deactivated with the use of dip 
switch #8.  The alarm speaker will sound during minimum, maximum or 
probe missing alarm conditions. 

Deactivated  

#8 

On 

Activated  

#8 

Off 

Note: 

Dip switches 1 & 2 are not active on this recorder 

Calibration 

Your instrument was carefully tested and calibrated before leaving the 
factory.  Additional calibration is not required.  However, if you wish to 
calibrate the recorder yourself, follow these procedures. 
1. Identify a standard that is more accurate than this unit. 
2. Place the probe of the unit and the standard into a controlled 

environment, allowing the unit to completely stabilize.  For best 
results, calibrate the unit at levels typically monitored during normal 
operation. 

3. To activate the calibration mode, turn the unit off.  Now press the 

ON/OFF key and the ALARM key at the same time.  The unit is in the 
“User Calibration” mode when the red LED is on.  The display will 
show a C during the warm-up period and then will flash between 0 
and F or 0 and C (depending on the measurement selected).  Match 
the reading of the recorder to the reading of the reference standard. 

4. To raise the temperature press the TIME (Up Arrow) key.  To lower 

the temperature, press the HOME (Down Arrow) key. 

To ensure you have matched the standard allow the pen to stabilize for 

at least 30 seconds before exiting calibration mode. 

5. When calibration is complete simply press the ON/OFF key to save 

the calibration settings. 

Calibration is stored in memory even after you turn the unit off.  User 

calibration information will not be lost if AC power fails. 

Chart Installation 

Use the following procedure to install or replace a chart in the recorder. 
1. Open the door by pressing the door latch. 
2. Press the HOME key to move the pen to the pen lift position.  The 

pen is automatically raised off the chart. 

3. Remove the old chart, place the appropriate new chart on the chart 

hub - being certain that the edge of the chart slides under the chart 
guide clips located at the outside of the chart. 

4. Press the HOME key to move the pen onto the chart. 
5. Set the time by pressing the TIME key, which rotates the chart.  The 

chart will continue to rotate until the key is released.  Release the 
TIME key when the appropriate time is indicated at the time indicator 
clip. 

Pen Installation 

A disposable fiber-tipped pen is installed in the recorder, when pen 
replacement becomes necessary, use the following procedure. 
1. Press the PEN HOME key to return the pen to the home position. 
2. Gently lift the pen and slide it off the pen arm. 
3. Slide the new pen onto the pen arm until the notch in the pen is 

secure to the pen arm. 

4. Remove the pen cap.  Insert it on the pen cap holder

 

located on the 

dial plate of the recorder. 

5. Press the PEN HOME key to return the pen to the recording position. 

Note:

 To prolong pen life, replace pen cap when recorder is not in 

use.  

Trouble Shooting 

Symptom Cause  Check/Remedy 

Pen trace to fine or 
absent. 

Tip too sharp, 
cartridge dried 
out. 

Sand tip, moisten or 
replace pen. 

Readings of pen and 
display do not match 
exactly. 

Wrong chart; Pen 
not on properly; 
Line thickness 

Check Dip Switch 4, 5 & 6; 
Push pen completely onto 
pen arm;  Verify that 
proper chart is installed. 

Out of Calibration or 
Questionable 
Accuracy 

Harsh 
environments 

See calibration procedure 
in this manual. 

Und (display)

 

Below Range of 
recorder 

Check temperature range. 

Ovr (display)

 

Above range of 
recorder 

Check temperature range. 

Prb (display) 

Probe missing 

Plug K-thermocouple into 
back of the recorder. 

Replacement Parts 

Item Part 

Number 

7- Day Chart, -50 to +50

°

F/

°

SPN 103194 

7- Day Chart, 0 to 50

°

F/

°

SPN 102989 

Pens, 6/Pack, Red 

13-940-997 

 
 