Stemi 1000, Stemi 2000, Stemi 2000-C
B 40-002 e 12/98
4-3
(2) Checking of the illuminating devices
Check the illuminating devices including a check of the power supply units used (switch settings, mains
connection) and if necessary, replace the halogen lamp.
•
Isolate the lighting unit (4-1/1) from the mains.
•
Undo the lamp housing (4-1/4) from the lamp
holder (4-1/2) by turning it lightly to the left.
•
Pull the defective halogen lamp (4-1/3) out of
the lamp holder and replace it by a new halogen
lamp of the same type (6 V 10 W).
IMPORTANT Do not touch lamp bulbs with your
bare hands; if necessary, clean the
bulb before the first use with pure
alcohol to prevent baking in of
soiling.
(3) Service
All tampering on optical parts or motion elements in the interior of the unit or on the power supply may
only be carried out by service specialists or specially authorised personnel.
For servicing, contact your nearest regional representative or
Carl Zeiss Jena GmbH
Zeiss Gruppe
Unternehmensbereich Mikroskopie
Tatzendpromenade 1a
D-07745 Jena
Telefon:
(03641) 64-2936
Telefax:
(03641) 64-3144
Internet:
micro@zeiss,de
http://www.zeiss.de
Figure 4-1
Lamp replacement
2
1
3
4