5.1 Visual inspection of probe tips
Although the probe tip is produced from wear-resistant material, it can become
damaged during the measurement. The probe tip wears over time and should
be replaced after approx. 20,000 measurements. To prevent faulty measure-
ments, the probe tips should therefore be visually inspected at regular intervals.
When to perform
•
Our recommendation: At the start of each series of measurements
•
Before each calibration
Required materials
•
Gage
•
Magnifying glass
Visual inspection of probe tips - procedure
►
Examine the probe tip under a magnifying glass.
What you can do next
•
Take measurements, see page 22
•
Run a calibration, see page 24
•
Replace the faulty probe tip, see page 36
Measuring probe with undamaged probe
tip
Example of a damaged probe tip (broken
tip)