24
FISCHERSCOPE
®
X-RAY
Set up
3.4
Establishing Connections
Check the supply voltage before connecting the instrument to the voltage supply. The nominal line
voltage is shown on the name plate and is either 110 V or 230 V (AC).
Figure 3.1: Connectors and switches at the right side of the instrument
Establish the following connections
1. USB: The USB cable has two different connectors: a quadratic one and a flat one. Connect the
quadratic connector to the instrument and the flat connector to a USB 2 socket at the PC. If no
free USB 2 socket is available, connect the mouse or the keyboard to a USB 3 socket.
The following table helps to distinguish between USB 2 and USB 3 sockets. Please note that not
all characteristics need to apply, e.g. a USB 3 socket may also be black.
ATTENTION
Wrong line voltage
Connecting the instrument to the wrong voltage can damage the instrument.
►
Make sure that the voltage, shown on the name plate of the instrument,
equals to your line voltage.
ATTENTION
Communication Breakdown Instrument – PC
If you connect the USB cable of the instrument with a USB 3 socket at the PC,
the communication between Instrument and WinFTM can be disturbed.
►
Connect the USB cable of the instrument with a USB 2 socket and not with
a USB 3 socket at the PC.
1
2
3
4
5
1 Main switch
2 Video signal line socket, to PC vid-
eo socket
3 USB interface, to PC
4 Main fuse
5 Voltage supply for instrument
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......