background image

5

11. Specifications

General

Measuring ranges

0...0.25 to 0...4.0 mbar (see 13. Ordering Code)

Nominal pressure

0.6 bar

Max. static pressure

overpressure safe up to permitted nominal pressure

Linearity

1% FS

Hysteresis

0.1% FS

Temperature drift

0.5% FS / 10K

Perm. ambient temperature

-10°...60°C

Perm. medium temperature

-20°...70°C

Protection class

IP 54 per DIN EN 60529

Electrical

Electrical connections

3-wire

3-wire

2-wire

Operating voltages

24 V DC

24 V DC

24 V DC

Power consumption

approx. 3 VA

approx. 3 VA

approx. 3 VA

Output signal

0-20 mA

4-20 mA

4-20 mA

Apparent ohmic resistance

700 

700 

max. 450 

Current limit

approx. 30 mA

approx. 30 mA

approx. 30 mA

Square-rooted output

± 0.5%

with suppression of leak flow volume

2% set up

Span adjustment

approx. 10% FS

Zero point adjustment

approx. 10% FS

Measuring Indication /

Switching Sections

Measuring indication

3½ digit LC-Display

Switching point adjustment

The digital display can be switched over between the differential pressure 
actual value and the switch point adjustments via selector. The display then 
indicates the related setpoint value. Setpoint values are adjustable within full 
scale range.

Switching point hysteresis

approx. 2%

Contact output

1 or 2 potential-free change-over contacts

Load data of contacts

U~

max.

 = 250 V AC, I~

max.

 = 2 A, P~

max.

 = 250 VA ohmic resistance

U=

max.

 =   30 V, I=

max.

 = 2 A, P=

max.

 = 60 W ohmic resistance

Connection

Electrical Connections

internal connector bloc, M16x1.5 connection
M12 plug connection 

Pressure Connections

female thread G¼, threaded hose coupling (al) for 6/8 mm flexible tube,
cutting ring connection (brass) for 6/8/10 mm tube

Materials

Housing

diecasted aluminium, varnished

Housing cover

ABS, self-extinguishing

Measuring element

silicium, glass, aluminium

Installation

wallmounting

vertical, pressure ports downward

any other orientation

correction of zero point recommended

Summary of Contents for DE61

Page 1: ...instructions use of this pro duct for purposes for which it is not intended and or operation of this product outside the limits spe cified for any of its technical parameters can result in harm to per...

Page 2: ...ssure safe shock and vibration resistant and long time stable The differential pressure acts on the sensor element and effects a measuring signal The electronic amplifier then generates an electrical...

Page 3: ...mA Correct offset using zero point potentio meter N Mount housing cover 5 4 Switching Point Adjustment The instrument s main board has two calibration poten tiometers and a coding switch to set switch...

Page 4: ...ust be read thoroughly to ensure that all of them work correctly when connected together 7 Transport The product must be protected against shock and vibra tion during transport It must therefore be pr...

Page 5: ...Measuring indication 3 digit LC Display Switching point adjustment The digital display can be switched over between the differential pressure actual value and the switch point adjustments via selector...

Page 6: ...6 12 Dimensions all units in mm unless stated otherwise Mounting holes 4 5 mm Cable gland M16 x 1 5...

Page 7: ...0 Screw connection AI for 6 mm flexible tube 4 0 Screw connection AI for 8 mm flexible tube 4 1 Electrical Output Signal 0 20 mA linear 3 wire A 4 20 mA linear 2 wire only 24 V DC without contacts an...

Page 8: ...ange without notice Changements techniques sous r serve Fischer Mess und Regeltechnik GmbH Bielefelder Str 37a D 32107 Bad Salzuflen Tel 49 5222 9740 Fax 49 5222 7170 eMail info fischermesstechnik de...

Reviews: