5
11. Specifications
General
Measuring ranges
0...0.25 to 0...4.0 mbar (see 13. Ordering Code)
Nominal pressure
0.6 bar
Max. static pressure
overpressure safe up to permitted nominal pressure
Linearity
1% FS
Hysteresis
0.1% FS
Temperature drift
0.5% FS / 10K
Perm. ambient temperature
-10°...60°C
Perm. medium temperature
-20°...70°C
Protection class
IP 54 per DIN EN 60529
Electrical
Electrical connections
3-wire
3-wire
2-wire
Operating voltages
24 V DC
24 V DC
24 V DC
Power consumption
approx. 3 VA
approx. 3 VA
approx. 3 VA
Output signal
0-20 mA
4-20 mA
4-20 mA
Apparent ohmic resistance
700
Ω
700
Ω
max. 450
Ω
Current limit
approx. 30 mA
approx. 30 mA
approx. 30 mA
Square-rooted output
± 0.5%
with suppression of leak flow volume
2% set up
Span adjustment
approx. 10% FS
Zero point adjustment
approx. 10% FS
Measuring Indication /
Switching Sections
Measuring indication
3½ digit LC-Display
Switching point adjustment
The digital display can be switched over between the differential pressure
actual value and the switch point adjustments via selector. The display then
indicates the related setpoint value. Setpoint values are adjustable within full
scale range.
Switching point hysteresis
approx. 2%
Contact output
1 or 2 potential-free change-over contacts
Load data of contacts
U~
max.
= 250 V AC, I~
max.
= 2 A, P~
max.
= 250 VA ohmic resistance
U=
max.
= 30 V, I=
max.
= 2 A, P=
max.
= 60 W ohmic resistance
Connection
Electrical Connections
internal connector bloc, M16x1.5 connection
M12 plug connection
Pressure Connections
female thread G¼, threaded hose coupling (al) for 6/8 mm flexible tube,
cutting ring connection (brass) for 6/8/10 mm tube
Materials
Housing
diecasted aluminium, varnished
Housing cover
ABS, self-extinguishing
Measuring element
silicium, glass, aluminium
Installation
wallmounting
vertical, pressure ports downward
any other orientation
correction of zero point recommended