Please familiarize all operating personnel with the general safety information in this summary.
Specific warnings and cautions are also found throughout this manual.
Warning
A warning statement in this manual identifies a condition or practice which if
not corrected or discontinued immediately, could lead to injury, illness, or death.
Do not use in the presence of flammable anesthetics.
This product has no user serviceable/replaceable parts. Please refer all
service to authorized service personnel.
This instrument is not intended to be used in the eye.
This product generates and records reference images from inputs obtained
from the software application FireflyPro. These reference images are not
intended for medical or other diagnostic purposes.
Caution
A caution statement in this manual identifies condition or practice, which if not
corrected or discontinued immediately, could lead to equipment failure,
equipment damage, or data loss.
In order to obtain optimum performance and ensure patient safety,
thoroughly read this manual.
Do not use the Firefly endoscope camera if you notice any signs of damage
to the components of the system. Contact customer service for immediate
assistance.
Do not attempt to disinfect the Firefly endoscope camera using
glutaraldehyde products, ethylene oxide gas, steam or any other liquid or gas
disinfectant.
The signal output (USB Port) connector is intended for connection of only
devices complying with IEC 60601-1, or other IEC standards (for example,
IEC 60950), as appropriate to the device. Connecting additional devices to
the Firefly endoscope camera is not allowed. You are responsible for
verifying that the system complies with the requirements of the system
standard IEC 60601-1-1.