Finish Thompson GP11 Series Installation And Maintenance Instructions Manual Download Page 4

3.  Install the rotating faces and cooling collar to the motor 

adapter.

 

a.   Carefully slide the housing cover over the impeller’s  

 

  shaft.  The side of the housing cover with the pressed-    

  in seal face should be away from the impeller.

 

b.   Separate the rotating head of the seal from the spring. 

 

NOTE:  If using a special seal material, it is vital that the 

special seal face mates against the housing cover’s station-

ary face.  

 

c.   Lubricate the impeller’s shaft, then carefully slide the    

  rotating face of the seal over the Impeller’s shaft using    

  a twisting motion. The carbon or special side of the  

 

  rotating face should be toward and pressed against  

 

  the stationary face of the housing cover.

 

CAUTION:  The carbon or special portion of the seal is 

easily damaged. Take care not to apply uneven force or  

crack the carbon or special material while installing.   

 

d.   Place the seal’s spring into position on top of the  

 

  installed rotating face.

 

e.   Relubricate the impeller’s shaft and then carefully slide   

  the second rotating face of the seal over the Impeller’s    

  shaft using a twisting motion (the carbon side of the    

  second rotating face should be facing away from  

 

  the other rotating face).  This will require some com-   

  pression of the seal’s spring.

 

f.   Place the cooling collar’s o-ring into position.  Slide the   

  cooling collar onto the Impeller’s shaft, with the  

 

  stationary face in the cooling collar pressed against  

 

  the second rotating face.  Continue to slide the cooling   

  collar until it mates against the housing cover.   Hold    

  the cooling collar in this position.  Take care to align    

  the water connections so they are accessible.

 

  NOTES:  

 

  1) The seal’s spring must be compressed further in  

 

  order to mate the cooling collar and housing cover.

 

  2) Take care not to pinch the cooling collar’s o-ring.  If    

  the o-ring is not properly seated, the cooling collar  

 

  will leak.

4.  Place the pump components onto the motor adapter.  

While holding the cooling collar against the housing cover, 

install the assembly into the motor adapter.  The motor 

adapter will hold the cooling collar in place.  It will now 

be necessary to hold the impeller to keep the seal’s spring 

compressed.

5.  Carefully place the assembly in position to the motor’s 

face.

 

a.  Screw the impeller onto the motor shaft or shaft adapter.   

  Use a drop of Blue 242 Loctite on the motor shaft or 

  shaft adapter (item 20) threads. It will be necessary to    

  hold the motor’s fan with a screwdriver to keep the motor  

  shaft from turning.

 

b.  Align the motor adapter so that the access hole is  

 

  straight up.

 

c.  Bolt the motor adapter to the motor using the hex-head   

  cap screws and lock-washers.

 

d.  Double check the alignment of the water connection    

  holes in the cooling collar.  

6.  Install the impeller housing.
 

a.  Install the housing cover’s o-ring.

 

b.  Place the impeller housing into position and install the   

  hardware. Tighten the six hex-head cap screws to 90    

 inch-pounds.

 

c.  Look through the discharge of the pump to verify prop   

  er spacing between the impeller and the housing cover.   

  If using a 56C or metric motor adapter, adjust the shaft   

  adapter if necessary (refer to the “assembly”section of    

  these instructions).

TROUBLESHOOTING

NO OR INSUFFICIENT FLOW

1.  Pump not primed.
2.  Closed valve.
3.  Viscosity too high.
4.  Air leaks in suction piping.
5.  Discharge head higher than anticipated.
6.  Suction lift too high or insufficient NPSH.  Check also for clogged 

suction line.

7.  Motor wired for wrong rotation.

INSUFFICIENT PRESSURE

1.  Air or gasses in liquid.
2.  Impeller diameter too small.
3.  Discharge head higher than anticipated.
4.  Motor wired for wrong rotation.

LOSS OF PRIME

1.  Leaking suction line.
2.  Foot valve or suction opening  not submerged enough.
3.  Foot valve too small or leaking.
4.  Air or gasses in liquid.
5.  Foreign matter in impeller.

EXCESSIVE POWER CONSUMPTION

1.  Head lower than rating.  Excessive flow.
2.  Specific gravity or viscosity of liquid is too high.

EXCESSIVE VIBRATION

1.  Loose piping or bolts.
2.  Pump cavitating from improper suction or feed.

    4

Summary of Contents for GP11 Series

Page 1: ...e four set screws from the shaft adapter onto the motor shaft d Remove the two temporary hex head cap screws and shim washers 4 Pull the motor adapter flush against the motor face Install GP11 SERIES Sealed Non Metallic Centrifugal Pumps Installation and Maintenance Instructions four hex head cap screws and flat washers through the motor adapter and into the motor and tighten 5 Verify there is a 0...

Page 2: ...n Refer to the directional arrow on the pump if needed NOTE A pump running backward will run but at a greatly reduced flow and pressure OPERATION FLOODED SUCTION SYSTEMS 1 Fully open the suction and discharge valves 2 If a cooling collar is used turn on the cooling water 3 Start the pump and verify liquid is flowing If there is no liquid flow refer to the Troubleshooting section of these instructi...

Page 3: ...with a rubber lubricant emulsion or soapy water b Carefully slide the housing cover over the Impeller s shaft The side of the housing cover with the pressed in seal face should be away from the impeller c Separate the rotating head of the seal from the large spring Relubricate the impeller s shaft then carefully slide the rotating head of the seal over the Impeller s shaft using a twisting motion ...

Page 4: ...tor adapter While holding the cooling collar against the housing cover install the assembly into the motor adapter The motor adapter will hold the cooling collar in place It will now be necessary to hold the impeller to keep the seal s spring compressed 5 Carefully place the assembly in position to the motor s face a Screw the impeller onto the motor shaft or shaft adapter Use a drop of Blue 242 L...

Page 5: ... RH V J100110 8 Retaining Ring Stainless steel J100111 9 Hex Bolt Stainless steel 6 required J100112 10 Flat Washer Stainless steel 6 required J100113 11 Hex Bolt NEMA 56C stainless steel 4 required J100114 IEC 80 frame stainless steel 4 required J103780 IEC 90 frame stainless steel 4 required J103662 Item Description Part No 12 Lock Washer NEMA 56C stainless steel 4 required J100115 IEC 80 frame ...

Page 6: ...to in whole or in part accident overload abuse chemical attack tampering or alteration The warranty does not apply to any other equipment used or purchased in combination with this product The manufacturer accepts no responsibility for product dam age or personal injuries sustained when the product is modified in any way If this warranty does not apply the purchaser shall bear all cost for labor m...

Page 7: ...is valid provided that the devices are fully assembled and no modifications are made to these devices Type of Device Centrifugal Pumps Models AC AK AV 400 500 600 800 GP 11 22 32 VKC 5 5 6 6H 7 8 10 DB 3 4 5 5 5 6 6H 7 8 9 10 11 15 22 MSKC SP 10 11 15 22 KC 3 4 5 5 5 6 6H 8 10 11 22 32 MSVKC UC 1516 1516L 1518 1518L 2110 3158 326 326H 328 436 438 4310H 326H 4310H 6410 EU Directives Machinery Safet...

Reviews: