4.

International Safety Certificates

UL Safety Recognition

CE 

certification

CSA Safety Recognition

5.

Product Specifications

Model No. 

VS08 

Voltage/Frequency 

110V/60Hz 

220V/50Hz 

Power (W) 

560 

Max Vacuum (mmHg) 

67

Max Flow Rate (L/min) 

110 

90 

Horsepower(HP) 

3/4

 

Noise level (dB) 

61 

Net Weight (kg) 

11 

Dimension (L x W x H) 

35 x 21 x 26 cm 

    PS: 1 mmHg 

  0.1333kPa 

  1.333 mbar

 

6.

Precautions

Please  read  this  manual  carefully  before  installing  or  operating  this equipment.
The instrument  should not be modified in any way.  Any modification will void 
the warranty and may pose a safety hazard. We are not responsible for any injury 
or  damage caused  by  any  non-intended  purposes  and  modifying  the  instrument 
without authorization. 

A. Check the voltage specified on the name-plate and ensure it matches the line 

voltage in your location.   

B. The pump is not completely corrosion-proof. Please do not operate it with 

corrosive gases directly.   

C. Install the instrument in a clean, dust-free, and well-ventilated area under 

40

.   

D. Finetech air filters can prevent large amounts of liquid from flowing into the 

pump. If needed, please contact the supplier.   

If the liquid comes into the pump, please stop using and contact the distributor 

immediately for service.   

E. To avoid damaging the pump easily, please vent the pump before starting the 

pump. 

F. This pump has a thermal protection device that automatically shuts off the 

pump when it becomes overheated. Keep pump upright during operation. 

G. The temperature on the surface of the pump is very high after or during use. 

Please don’t touch the surface to avoid being burned.   