14 of 26 

Fluid Viscosity 

 

The portable analysers are designed to handle a fluid viscosity of up to 460 cSt for the small unit or 
2400 cSt for the larger units

. It’s possible to pass high viscosity oils through the unit if they are 

preheated as per the below table. 

Note that as a pre-heated oil passes though the unit it will cool, therefore increasing its viscosity and 
strain on the pump. 

Temperature 

°C 

Viscosity (cSt) 

ISO 32 

ISO 46 

ISO 68 

ISO 100  ISO 150  ISO 220  ISO 320  ISO 460  ISO 680 

10.0 

161 

262 

442 

711 

1185 

1921 

2911 

4827 

8042 

12.0 

141 

227 

379 

605 

1001 

1611 

2435 

3991 

6588 

14.0 

124 

198 

327 

518 

850 

1358 

2048 

3318 

5427 

16.0 

109 

173 

283 

446 

726 

1150 

1730 

2774 

4496 

18.0 

97 

152 

246 

385 

622 

979 

1469 

2331 

3745 

20.0 

86 

134 

215 

334 

536 

838 

1254 

1968 

3135 

22.0 

77 

118 

188 

291 

464 

720 

1074 

1670 

2638 

24.0 

69 

105 

166 

255 

403 

621 

925 

1423 

2230 

26.0 

62 

94 

147 

224 

352 

538 

799 

1219 

1894 

28.0 

56 

84 

130 

198 

308 

469 

694 

1048 

1616 

30.0 

51 

75 

116 

175 

271 

409 

604 

905 

1385 

32.0 

46 

68 

103 

155 

239 

359 

528 

784 

1192 

34.0 

42 

61 

93 

139 

212 

316 

464 

683 

1030 

36.0 

38 

56 

83 

124 

188 

279 

408 

596 

894 

38.0 

35 

50 

75 

111 

168 

247 

361 

523 

778 

40.0 

32 

46 

68 

100 

150 

220 

320 

460 

680 

42.0 

29 

42 

62 

90 

135 

196 

285 

406 

596 

44.0 

27 

38 

56 

82 

121 

176 

254 

360 

525 

46.0 

25 

35 

51 

74 

109 

158 

227 

320 

463 

48.0 

23 

33 

47 

67 

99 

142 

204 

285 

411 

50.0 

21 

30 

43 

62 

90 

128 

183 

255 

365 

 

Internal Gear Pump 

 
The internal gear pump provides a steady flow for accuracy of results. Fluids above the rated 
viscosity can cause excess motor strain and cause it to overheat or slip. 

It’s important to pay 

attention to the grade of oil being analysed and its temperature. 

 

Summary of Contents for FS9V4

Page 1: ...mail sales filtertechnik co uk Filtertechnik 1 Central Park Lenton Lane Nottingham NG7 2NR Registered in England No 03969985 VAT No 760 821 731 Certification No UK002838 1 of 26 FS9V4 Fuel Oil Cleanli...

Page 2: ...Unit 8 On Off Switches 8 Touch Screen Layout 8 Priming and Flushing the Pump 9 ISO Alarm Limits 10 Datalogging 10 Time Zone Adjustment 11 Entering a Test Name 12 Battery Charging 13 Battery Charging...

Page 3: ...2NR Registered in England No 03969985 VAT No 760 821 731 Certification No UK002838 3 of 26 RH Water Content 20 Water Sensor Saturation RH 21 Water Sensor Parts Per Million PPM 21 Density Viscosity Sen...

Page 4: ...eration Tank sampling Bottle sampling 200ml minimum sample bottle recommended Cleanliness standards ISO 4406 NAS 1638 SAE AS4059 Pump type Gear pump Duty cycle Continuous Viscosity range Small Unit 1...

Page 5: ...ne Nottingham NG7 2NR Registered in England No 03969985 VAT No 760 821 731 Certification No UK002838 5 of 26 Component Identification Key 1 Touch screen display 5 Return Outlet Minimess Test Point 2 P...

Page 6: ...pling is removed so the connections won t leak during transit To connect the hoses screw them onto the Minimess Test Points hand tight only Make sure the hose assembly with the strainer is always used...

Page 7: ...ith a strainer fitted to the suction tube assembly The inline strainer has a 270 micron mesh and is used for keeping accidental debris from damaging the pump gears or clogging other components The str...

Page 8: ...Units have two push button switches one for the pump and one for the touch screen Note that because the screen regulates the pump speed the pump will only work when the screen is switched on Touch Scr...

Page 9: ...d in England No 03969985 VAT No 760 821 731 Certification No UK002838 9 of 26 In the top right corner of the screen is a settings icon clicking this will open the settings page At the bottom left of t...

Page 10: ...he ISO code on the main screen will turn green when dirty it will become red between the two values the text is white Datalogging The unit contains two internal dataloggers one for the graph and one f...

Page 11: ...e Nottingham NG7 2NR Registered in England No 03969985 VAT No 760 821 731 Certification No UK002838 11 of 26 Time Zone Adjustment The display screen is pre set with the date and date to GMT 0 UK time...

Page 12: ...o UK002838 12 of 26 Entering a Test Name The main screen shows a Test Name field which can be clicked to bring up a new window Here a test name can be entered and when the back button is clicked this...

Page 13: ...for prolonged periods of time the battery is best left at 60 charge to maintain an optimum lifespan Battery Charging Unit The battery charge unit is designed for indoor use only and should not come in...

Page 14: ...24 198 327 518 850 1358 2048 3318 5427 16 0 109 173 283 446 726 1150 1730 2774 4496 18 0 97 152 246 385 622 979 1469 2331 3745 20 0 86 134 215 334 536 838 1254 1968 3135 22 0 77 118 188 291 464 720 10...

Page 15: ...range 0 100 A reading of 0 would indicate a fluid free of water while a reading of 100 would indicate a fluid that is saturated with water see diagram below The capacitance sensor absorbs water molec...

Page 16: ...ssure 0 5 to 50 bar Voltage supply 1 32 VDC Power consumption 30 60 mA Output 4 20 mA Protection type IP67 Recalibration Intervals depend on use recommended annually Sensor Working Range The graph sho...

Page 17: ...ith other fluid types will provide inaccurate readings Note If the fluid sample contains free water or has a water content of above 300ppm cloudy hazy appearance it should not be passed through the po...

Page 18: ...ings Specifications Specification Details Accuracy Density 2 Viscosity 5 Measurement Range Density 0 6 g cm 1 3 g cm Viscosity 25 400mPa s Resolution Density 0 001g cm Viscosity 0 1mPa s Working Volta...

Page 19: ...th a maximum allowable inlet pressure of 350 bar reducing this down to 2 5 bar at the outlet It has a working viscosity range of 1 300 cSt The High Pressure Device can be purchased separately and used...

Page 20: ...pull fluid through Check the hoses and strainer for blockages Screen is behaving erratically Dirt on the screen or hands Wipe the screen clean to remove traces of dirt or oil In most cases gloves will...

Page 21: ...ten its useful life Water Sensor Parts Per Million PPM Exposing the PPM sensor to diesel above 300ppm will cause sensor lockout This be resolved by passing clean dry diesel 100 ppm through the unit un...

Page 22: ...below shows the ISO 4406 codes by number of particles per 1 ml of fluid sampled ISO 4406 Code Counts mL Greater than Up to including 0 0 0 01 1 0 01 0 02 2 0 02 0 04 3 0 04 0 08 4 0 08 0 16 5 0 16 0 3...

Page 23: ...0 9 9500 25 9 304000 6 0 32 11 10 16 320 21 10000 26 320000 6 1 0 352 11 1 11 16 1 352 21 1 11000 26 1 352000 6 2 0 384 11 2 12 16 2 384 21 2 12000 26 2 384000 6 3 0 416 11 3 13 16 3 416 21 3 13000 26...

Page 24: ...00000 8 6 2 02 13 6 64 18 6 2020 23 6 64000 28 6 2020000 8 7 2 14 13 7 68 18 7 2140 23 7 68000 28 7 2140000 8 8 2 26 13 8 72 18 8 2260 23 8 72000 28 8 2260000 8 9 2 38 13 9 76 18 9 2380 23 9 76000 28...

Page 25: ...delivery A product claimed to be defective must be returned to the place of purchase Filtertechnik at its sole option shall replace the defective product with a comparable new product or repair the d...

Page 26: ...Representative s Name N A EC Representative s Address N A Equipment Description Portable analysers for fuel and oil Equipment Model Designation FS9V4 Application of Council Directive EMC Directive 20...

Reviews: