Fieldpiece SDMN6 Operator'S Manual Download Page 2

11

16

12

17

13

18

14

19

15

20

Pressure Switch Test

WARNING:

 Read "Pressure Switch Testing Safety 

Info" section below prior to testing.

1.  Connect leads (interchangeable) from the 

SDMN6 to the N.O. (normally open) and COM 

(common) terminals of the pressure switch. 

2. Connect the 2 yellow hoses of the Y-splitter 

to the P1 and PUMP ports of your SDMN6.  

3.  ZERO your SDMN6 while at ambient pressure 

with yellow hoses attached. For measure-

ments less than 2 inWC, take reading within 

1 minute after zeroing for best accuracy.

4. Press TEST to start the pump. A brief warm up 

period ensues. 

5. Now connect the white hose to the negative 

side of the pressure switch. 

(Incorrect setup 

will cause incorrect pressure measurements.)

6. Make sure the pressure switch is vertically 

oriented, the same as it is on the equipment.

7. 

If the CLOSED LED is OFF

, press   to increase 

the pump speed and induce a negative pres-

sure on the switch. The CLOSED LED will light 

when the pressure switch closes. Note the 

closing pressure. 

 

If the CLOSED LED is ON

, press   to slow the 

pump and decrease the negative pressure  on 

the switch. The CLOSED LED turns off when 

the switch opens.  Note the opening pressure.

NOTE:  For higher rated switches, you can hold 

down   or   to accelerate/decelerate the 

pump faster. Slow your acceleration as you 

approach the switch's rated pressure.

NOTE: A pressure switch should open and close 

within the manufacturer's tolerance rating of 

the switch (typically ±10%).

AUTO CAPTURE MODE

Press HOLD to automatically HOLD a closing 

or opening pressure of a switch as you change 

the speed of the pump. "HOLD" blinks until a 

change in continuity is detected. Press HOLD 

again to exit AUTO CAPTURE mode.

PUMP BLEED ACCESSORY

Most pressure switches have 

a bleed port to help regulate the 

pressure required to close and open 

the switch. 

Pressure switches rated around 

-0.10inWC often 

don't

 have a bleed 

port. You can insert the Pump Bleed 

Accessory directly onto the PUMP 

port to give the pump better control 

of these low negative pressures.

Indications a pressure switch is out of 

calibration, has a weak/bad diaphragm, or 

a bad micro switch in the pressure switch:

•   Closing/opening outside the manufacturer's 

tolerance rating.

•   CLOSED LED blinks rapidly.

•   Dramatic fluctuations of pressure readings.

Checking the Diaphragm

If you have difficulty maintaining the correct 

pressure to open and close the switch the 

diaphragm could be cracked.

Checking the Bleed Port

 Pressure readings are greatly affected when 

the hose to the pressure switch is pinched. If the 

pressure does not change when you pinch the 

hose on a pressure switch with a bleed port, the 

bleed port could be blocked.

Examine the bleed port for signs of corrosion 

or debris. Corrosion is a sign that the switch 

has been exposed to excessive moisture and 

diaphragm could be coated with debris.

Checking the Inducer

If the pressure created by the inducer 

measures more than enough to close the 

pressure switch, but the SDMN6 proves the 

pressure switch to be good, the inducer may not 

be providing enough airflow to close the switch.

Adjustable Pressure 

Switch Calibration

WARNING:

 Read "Pressure Switch Testing Safety 

Info" section below prior to testing. 

1. Follow steps 1-6 of "Pressure Switch Test".

7. Press   /    to adjust the pump speed until 

displayed negative pressure is stable, within 

±0.02 inWC of your target pressure.

8. Start with the switch closed. If the switch is 

open, slowly loosen the adjustment screw 

until the CLOSED LED turns on.

9. Open the switch by slowly tightening the 

adjustment screw on the adjustable pressure 

switch until the CLOSED LED turns off. The 

adjustable pressure switch is now calibrated 

to your target pressure.

NOTE: Stabilized pressures may slightly fluctu-

ate after an adjustment. Be sure to adjust the 

pump speed (   /   ) as needed to maintain 

the desired induced negative pressure.

NOTE: Dual-pressure switches only have one 

spring and one diaphragm so it's only neces-

sary to test the negative side.

Transducer Test

1.  Connect the 2 yellow hoses of the Y-splitter 

to the P1 and PUMP ports of your SDMN6. 

2. Connect a volt meter to the transducer's 

terminals.

3.  ZERO your SDMN6 while at ambient pressure 

with yellow hoses attached. For measure-

ments less than 2 inWC, take reading within 

1 minute after zeroing for best accuracy.

4. Press TEST to start the pump. A brief warm 

up period ensues. 

5. Now connect the white hose to the low side 

of the transducer. 

(Incorrect setup will cause 

incorrect pressure measurements.)

6. Press   /   to increase/decrease the pump 

speed and resulting negative pressure. For 

higher rated transducers, you can hold down 

the button to get the pump up to speed 

faster. 

7.  As you increase the negative pressure, the 

voltage across the terminals will change 

based on the transducer's rating. Compare 

observed voltage to rated voltage to deter-

mine the health of the transducer.

Pressure Switch 

Testing Safety Info

The pressure switch is a safety device that 

prevents the furnace from running in an unsafe 

condition. An unsafe condition can result in 

injury, loss of property or even loss of life.

1. Before every use, check the SDMN6 and the 

tubing for breaks or blockage. Check the tub-

ing for any moisture as well. 

2. It is critical that the hoses going to P1 and the 

switch are the same length.

3. Check for the presence of moisture in the 

pressure switch before using the SDMN6. Do 

not use the SDMN6 on a pressure switch that 

has moisture in it. Moisture can damage the 

SDMN6 and VOID the warranty. Check the 

tubing for moisture build-up as well.

4. The SDMN6 is designed to produce a precise 

amount of negative pressure. Always zero the 

SDMN6 before testing a switch. 

5. Fieldpiece is not responsible for incorrect 

readings by faulty test equipment or un-

trained personnel.

6. In the interest of safety, the SDMN6 should 

only be used by trained, competent profes-

sionals who understand the hazards and 

consider the risks of working on and with 

tools and equipment.

7. The pressure rating for the pressure switch is 

either stamped on the pressure switch or on 

a sticker attached to the pressure switch. Oth-

erwise, contact the vendor or manufacturer of 

the furnace to get that information.

8.  An adjustable pressure switch must open at 

the desired negative pressure to match the 

one it is replacing.

9.  Fieldpiece is not liable for any damages, in-

juries, or loss of life incurred by the misuse of 

the SDMN6.

NOTE: Pressure switch ratings are noted by  

“[pressure] PF”. This means that the circuit 

opens when the switch senses its rated Pres-

sure Fall.  Example:  A -0.40" WC PF rated 

switch should open at a negative 0.40 inches 

of water column. The pressure switch should 

close and more importantly “open” within the 

manufacturer’s tolerance rating.

 

     

!

  WARNINGS  

!

Turn off power to furnace prior to testing 

to avoid low voltage shorting. 

Limited Warranty

This meter is warranted against defects in 

material or workmanship for one year from 

date of purchase. Fieldpiece will replace or 

repair the defective unit, at its option, subject to 

verification of the defect. This warranty does not 

apply to defects resulting from abuse, neglect, 

accident, unauthorized repair, alteration, or 

unreasonable use of the instrument.

Any implied warranties arising from the sale of 

a Fieldpiece product, including but not limited 

to implied warranties of merchantability and 

fitness for a particular purpose, are limited to 

the above. Fieldpiece shall not be liable for loss 

of use of the instrument or other incidental or 

consequential damages, expenses, or economic 

loss, or for any claim of such damage, expenses, 

or economic loss. State laws vary. The above 

limitations or exclusions may not apply to you.

For Service

In the USA, call Fieldpiece Instruments for 

one-price-fix-all out of warranty service pricing. 

Send check or money order for the amount 

quoted. Send the meter freight prepaid to 

Fieldpiece Instruments. Send proof of date and 

location of purchase for in-warranty service. 

The meter will be repaired or replaced, at the 

option of Fieldpiece, and returned via least 

cost transportation. Outside of the USA, please 

visit www.fieldpiece.com for service contact 

information.

For international customers, warranty for 

products purchased outside of the U.S. should 

be handled through local distributors.

www.fieldpiece.com

    ©Fieldpiece Instruments, Inc  2014; v14

PUMP

P1

P2

N.O.

COM

NEG

To Pump

Port

To Yellow 

Pump Hose

Reviews: