Fieldpiece SC54 Operator'S Manual Download Page 2

11

16

12

17

13

18

14

19

15

20

Amps AC (AAC) TrueRMS (50-60Hz)

Test any isolated power line.  Select AAC/Hz dial position.

Ranges:

  0 to 400AAC  

Resolution:

  0.1A 

Crest factor:

  ≤ 3

Accuracy:

  ±(2.0% + 5dgts) 50-60Hz

Jaw Opening: 

1.2in 

(

30 mm)  

Overload Protection:

 400AAC 

Frequency (Hz) Through Clamp

Measure frequency without using test leads, just use the clamp. 

Turn dial to AAC/Hz and press SEL. Clamp Hz will be displayed.

Range:

  20Hz to 400Hz  

Accuracy:

  ±(0.5% + 5)

Minimum current range: 

 > 5AAC at 20 to 100Hz, >10AAC at 100 

to 400Hz  

Resolution:

  0.1Hz 

Overload Protection: 

400AAC

Resistance (Ω)

Used for “ohming out” a motor.  0.1Ω resolution is necessary to 

test the resistance between the motor poles because the values are 

typically very low.  Select MFD/Ω/

O

F

F

MFD

NCV

V

AC

V

DC

A

AC

Hz

Hz

%

°C

°F

SC56

AUTO OFF

TRUE RMS

CAT III

600V

TEMP

K-TYPE

30V 

MAX

SEL

HOLD

INRUSH

 and press SEL button once.

Ranges:

  0 to 1000Ω, 1000Ω to 9999Ω 

Resolution:

 0.1Ω 

Overload Protection:

  600VDC/VAC rms

Accuracy:

  ±(1.5% + 5) 

Continuity  (

O

F

F

MFD

NCV

V

AC

V

DC

A

AC

Hz

Hz

%

°C

°F

SC56

AUTO OFF

TRUE RMS

CAT III

600V

TEMP

K-TYPE

30V 

MAX

SEL

HOLD

INRUSH

)

Use the continuity feature to test if a circuit is open or closed.  Use 

this feature to check fuses as well.  A steady “beep” and green LED 

indicate the circuit is good. Select MFD/Ω/

O

F

F

MFD

NCV

V

AC

V

DC

A

AC

Hz

Hz

%

°C

°F

SC56

AUTO OFF

TRUE RMS

CAT III

600V

TEMP

K-TYPE

30V 

MAX

SEL

HOLD

INRUSH

 and press SEL twice. 

Range:

 1000Ω  

Resolution:

 0.1Ω  

Response time:

 100ms

Audible beep:

 <30Ω 

Overload Protection:

 600VDC/VACrms

Non Contact Voltage (NCV)

Use the non contact voltage (NCV) feature to test if a wire is hot 

or not.  Always test on a known live source before using. A red LED 

blinks and beeping sound is emitted at voltages >24VAC

AC Voltage Detection Range:

  24VAC to 600VAC (50-60Hz)

Temperature (°F/°C)

Plug any K-type thermocouple directly into the meter to measure 

temperature. Cold junction is located inside the meter and allows for 

extremely accurate measurements even in rapidly changing ambient 

temperatures (going from rooftop to freezer). No adapter is required.  

See Temp Calibration section for calibration instructions.  

Range:

 -40°F to 2200°F, (-40°C to 1200°C)  

Resolution:

  0.1°

Accuracy:

    ±(1°F) 32°F to 120°F,  ±(1°C) 0°C to 49°C

±(1%+2°F) 120°F to 750°F,  ±(1%+1°C) 49°C to 400°C

±(2%+6°F) -40°F to 32°F,  ±(2%+3°C) -40°C to 0°C

±(2%+6°F) 750°F to 2200°F,  ±(2%+3°C) 400°C to 1200°C

Sensor type:  

K-type thermocouple

Overload protection:  

30 VDC or 30VAC rms

Temp. Calibration

For accuracies of ±1°F, 

calibrate to a known 

temperature.  A glass 

of stabilized ice water 

is very close to 32°F 

(0°C) and is usually very 

convenient but any 

known temperature can 

be used.  

1.  Select the °F/ °C range.

2.   Plug thermocouple to be 

calibrated into the K-type jack.

3.  Unscrew A and B and remove 

the battery cover.

4.  Stabilize a large cup of ice 

water. Stir the ice with the 

water until temperature stays at 

32°F (0°C).

5.  Immerse the thermocouple 

probe and let it stabilize. Keep 

stirring to prevent micro-

environments.

6.  Use a small screwdriver to 

adjust calibration pot C to the 

right of the battery as close to 

32°F as you would like. 

Modular 

Expandability

Your Swivel Clamp Meter is compatible with 

all Fieldpiece Accessory Heads. With Fieldpiece 

Accessory Heads,  you can measure any 

available parameter, and read the measurement 

on your meter's display in real-time. 

Just set the range to mVDC. Pressing SEL 

changes the mV range. Remove the probe tips 

of your test leads, and connect your accessory 

head (model AAV3 shown). 

Visit www.fieldpiece.com to see all of the 

different accessory heads that Fieldpiece offers. 

Safety Features

1.  Bright LED and beeper warn you when testing 

voltages above 30V.

2.  Switch to the NCV function (non contact volt-

age) and point clamp claw towards suspected 

voltage source. Monitor the bright LED and 

beeper to see if the source is “hot.”

3. Rotate the clamp to the angle that best suits 

the situation.

4.   Temp switch to prevent leaving thermocouple 

plugged in while measuring voltage.

BatteryReplacement

The battery must be replaced when the 

battery icon is empty. "bAtt" will display with a 

beeping sound. Meter will shut off in 5 seconds.

Disconnect and unplug leads, turn meter off 

and remove battery cover.    

Safety Information

•   Never ground yourself when taking electrical 

measurements.  Do not touch exposed metal 

pipes, outlets, fixtures, etc., which might be at 

ground potential, while taking measurements.  

Keep your body isolated from ground by us-

ing dry clothing, rubber shoes, rubber mats, 

or any approved insulating material.

•   Disconnect the test leads before opening the 

case. Inspect the test leads for damage to the 

insulation or exposed wire.  Replace if suspect.  

Keep your fingers behind the finger guards on 

the probes while taking measurements.

•   When disconnecting from a circuit, disconnect 

the “RED” lead first, then the common “BLACK” 

lead.  Use one handed testing when possible.  

Work with others.

•   Turn off power to the circuit under test before 

cutting, unsoldering, or breaking the circuit.

•   Do not measure resistance (ohms)  when cir-

cuit is powered. Isolate load by disconnecting 

from circuit.

•   Disconnect the meter from the circuit before 

turning any inductor off, including motors, 

transformers, and solenoids.  High voltage 

transients can damage the meter beyond 

repair.  Do not use during electrical storms.

•   Do not apply more than rated voltages be-

tween input and ground.

•   Isolate capacitors from system and discharge 

them safely before testing.

All voltage tests:  

All voltage ranges will 

withstand up to 600V.  Do not apply more than 

600VDC or AC rms.

Symbols used:

 

Caution, risk of electric shock

 

Caution, refer to manual.

 Ground

 

Double insulation

Maintenance

Clean the exterior with a dry cloth.  Do not 

use liquid.

Limited Warranty

This meter is warranted against defects in 

material or workmanship for one year from 

date of purchase. Fieldpiece will replace or 

repair the defective unit, at its option, subject 

to verification of the defect.

This warranty does not apply to defects 

resulting from abuse, neglect, accident, 

unauthorized repair, alteration, or unreasonable 

use of the instrument.

Any implied warranties arising from the sale of 

a Fieldpiece product, including but not limited 

to implied warranties of merchantability and 

fitness for a particular purpose, are limited to 

the above. Fieldpiece shall not be liable for loss 

of use of the instrument or other incidental or 

consequential damages, expenses, or economic 

loss, or for any claim of such damage, expenses, 

or economic loss. 

State laws vary. The above limitations or 

exclusions may not apply to you.

Obtaining Service

Call Fieldpiece Instruments for one-price-

fix-all out-of-warranty service pricing. Send 

check or money order for the amount quoted. 

Send the meter freight prepaid to Fieldpiece 

Instruments. Send proof of date and location 

of purchase for in-warranty service. The meter 

will be repaired or replaced, at the option 

of Fieldpiece, and returned via least cost 

transportation.

www.fieldpiece.com

© Fieldpiece Instruments, Inc  2012;  v04

A

B

C

O

F

F

O

F

F

SE

L

RN

G

O

F

F

SE

L

RN

G

AUTO OFF

TRUE RMS

CAT III

600V

TEMP

K-TYPE

30V 

MAX

Unplug Leads and Slide 

TEMP Switch to the Right

O

F

F

!

  WARNINGS

DISCONNECT AND UNPLUG TEST LEADS before opening case.

TEST NCV FUNCTION ON KNOWN LIVE WIRE before using.

DO NOT APPLY VOLTAGE greater than 30VAC or 60VDC to the 

thermocouple or the jacks when the rotary dial is on °F. (Use 

only k-type thermocouples)

REMOVE THE THERMOCOUPLE when taking voltage measurements.

DISCONNECT TEST LEADS when measuring temperature.

DO NOT APPLY VOLTAGE TO THE JACKS when the rotary dial is on 

microamps. Even low voltages can cause a current overload and 

potentially harm the meter.

!

SEL

RNG

Air Velocity

& Temperature

Head

English

READ

LO BATT

ON

LCD X
100

Average

(16 sec)

Metric

Real

time

AAV3

AUTO-

OFF

English

Metric

Ft/min

M/s

KM/hr

MPH

ºF

ºC

OFF

O

F

F