5.4 Troubleshooting Procedures
Section 5: Troubleshooting
CV-6SLX User Manual
5-20
0101-8242-0, Rev. C
DANGER: HIGH VOLTAGE
Before performing any of the procedures described on this page, perform the
procedure described in section 5.4.1.
Table 5-8 No Beam Emitted After HV and Gun Are Switched On
Symptom: HV and gun are switched on, but no beam is emitted
For additional information about this fault mode, see Ishikawa Diagram #3.
Possible Causes
Procedure
1. Faulty or loose connections.
1.
Make sure that all HV and FPS connections are correctly and securely made.
2. Confirm HV Ready signal
2.
See Table 5-3.
3. Confirm Gun Ready signal
3.
See Table 5-4.
4. Emitter damaged
4.
Follow instructions in gun manual to troubleshoot emitter issues.
5. Confirm HV IS ON signal
5.
See Table 5-5.
6. Confirm GUN IS ON signal
6.
See Table 5-6.
7.
Fault conditions whose investigation
requires removal of FPS top cover.
7.
Contact Temescal Field Service
.
DANGER: HIGH VOLTAGE
Before performing the following procedure, perform the procedure described in
section 5.4.1.
Table 5-9 HV Output Unstable
Symptom: HV fluctuates or sags from requested level
For additional information about this fault mode, see Ishikawa Diagram #4.
Possible Causes
Procedure
1. Incorrect or fluctuating line voltage
1.
Confirm input line voltage
2. Faulty or loose cable connections.
2.
Make sure that all power module and FPS cable connections are
correctly and securely made.
3. Emitter damaged
3.
Follow instructions in gun manual to troubleshoot emitter issues.
4. System incorrectly grounded for HV
4.
Make sure that grounding instructions in e-gun manual have been
correctly implemented.
If the preceding steps have failed to correct the fault, perform the following procedure:
1. Switch off the power supply.
2. Disconnect HV output cable.
3. Switch on the power supply and the HV and leave them switched on for no more than 9 seconds, observing whether the
symptom persists during that interval.
CAUTION
DO NOT leave the high voltage on for more than 9 seconds with the HV cable disconnected. Doing so will
result in the emitter’s filament burning out.