Perform Test 

 

 

If the display is blank, press the operate button 

once. The display will show the last test results. 

 

1.   Stand up.  Hold the eSense

TM 

Piko

®

 in 

your hand.  Do not cover the vent holes. 

   

2. 

Bring the eSense

TM 

Piko

®

 up with the 

mouthpiece towards your mouth. 

 

3. 

Press the operate button.  A short beep 

will confirm the action.  

 

4. 

Wait to hear the second beep.  The blow 

animation will indicate that the test can 

begin.   

 

5. 

Inhale as fully as you can.   

 

6. 

Place the mouthpiece in your mouth and 

blow as hard as you can for several 

seconds. 

 

7. 

Your test results are displayed as soon as 

you complete the test. 

 

8. 

If a reference value was set, a zone indicator 

will show your color zone.   

 

9. 

You may want to perform 3 Tests. 

 
10.  The 

eSense

TM 

Piko

®

 

will select and save 

only the best results for any tests performed 

within 3 minutes. 

 

 

Summary of Contents for eSense PiKo

Page 1: ...USER S MANUAL eSense PiKo Ferraris Respiratory Inc 901 Front Street Louisville CO 80027 USA Tel 303 666 5555 Fax 303 666 5588 Website www ferrarisrespiratory com ...

Page 2: ...ing two conditions 1 this device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation Changes or modifications not expressly approved by Ferraris Respiratory Inc could void the user s authority to operate the eSenseTM Piko ...

Page 3: ...ative to the patient s PEF or FEV1 reference values In addition the blow quality is evaluated to notify the patient to repeat the test in case of a cough or any suspect blow PEF the fastest speed a person can blow air out of their lungs FEV1 the amount of air blown out in the first second of a forced exhalation Color Zones display the test results in a traffic light color format relative to the pa...

Page 4: ...igns or symptoms of chest tightness shodness of breath coughing or wheezing occur contact your healthcare professional immediately The eSenseTM Piko is intended for use by a single person Follow instructions carefully to get a correct reading If you use the eSenseTM Piko above 1 000 feet elevation 300m adjust your values by adding 1 5 To do this multiply the PEF FEV1 by 1 015 The eSenseTM Piko is ...

Page 5: ...ata Liter per minute 10 PEF indicator 11 Memory counter 12 FEV1 indicator 13 FEV1 data Liters 14 Battery cover behind Note The PEF data items 9 10 and the FEV1 data items 12 13 alternate at 2 second intervals Note The display turns blank if the operate button is not pressed for 3 minutes Upon pressing once the operate button the display will show the last test results Front View Back View ...

Page 6: ...confirm the action 4 Wait to hear the second beep The blow animation will indicate that the test can begin 5 Inhale as fully as you can 6 Place the mouthpiece in your mouth and blow as hard as you can for several seconds 7 Your test results are displayed as soon as you complete the test 8 If a reference value was set a zone indicator will show your color zone 9 You may want to perform 3 Tests 10 T...

Page 7: ...will show briefly the new reference value and return to show the last test results Notes During the reference setting process if the button is not pressed within 5 seconds the unit will exit the Setting Mode without modifying the old reference The display will return to the last test results The color zone and relative warning indicator in the memory review mode will reflect the new reference valu...

Page 8: ...ter location 2 Release the button Each press of the button will now display the next oldest stored test 3 There are 2 ways to exit the Memory Review Mode Press and hold down the Operate button for 5 seconds Avoid pressing the Operate button for 20 seconds The unit will then display the latest test result and will be ready for a new test Note The entire memory bank can be erased by setting the PEF ...

Page 9: ...ons Default Settings G Green test results over 80 of the set value Y Yellow test results between 50 80 of the set value R Red test results under 50 of the set value Note If your test result falls in the Red Zone contact your healthcare professional immediately Q Factor Indicator Indicates to repeat the test since one or more of the following occurred A cough was detected The blow effort was too sh...

Page 10: ...ds and other relevant information refer to the Manual of the optional program Note It is possible to interrupt the downloading process at any time by removing the eSenseTM Piko from the IR cradle The memory bank can also be downloaded to a LogPad Palm PDA via Personal Area Network using the built in transceiver Please see LogPad manual for instructions downloading eSenseTM Piko data Sound Patterns...

Page 11: ...llowed by rinsing Hold the unit upside down and shake to remove the water Wipe exterior with a clean dry towel Air dry completely before further use or storage 4 To avoid damage to the unit DO NOT immerse in water or use a dishwasher DO NOT use a high pressure water flow DO NOT use any sort of active solvent for cleaning DO NOT dry by heat hair dryer or a dishwasher DO NOT insert anything into the...

Page 12: ...on Bi Directional IR port RS232 format RF via proprietary protocol External settings Possible settings using the optional software Each Color Zone limits in I 0 increments Select Coior Zones to relate to PEF or FEV1 Battery 2 type 357 Silver Oxide button cells or equivalent Battery life 0 5 year based on average of 3 Tests day or approximately 5 400 consecutive Blows Dimensions 75 X 35 X 20 mm 3 X...

Page 13: ...n any way that in FRI s judgment affects its durability or reliability or any product or part subjected to misuse negligence or accident THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES OF MERCHANTABILITY AND FITNESS A copy of the receipt must accompany a warranty claim If the product should become defective within the warranty period return to FRI at 901 Front Street Louisville CO 80027 USA Safet...

Page 14: ...ction complies with the European Union Medical Device Directive This symbol indicates that this device provides a certain level of protection against electrical shock Ferraris Respiratory Inc 901 Front Street Louisville CO 80027 USA Tel 303 666 5555 Fax 303 666 5588 Website www ferrarisrespiratory com In Europe Hartforde Court John Tate Road Hertford SG13 NW UK ...