6- TECHNICAL SPECIFICATIONS

INDICATION

Display

5 Digits, 7 Segments , Red Led High Brightness

Digit Height

14,2 mm. / 0,56’’

Filter

Anti-reflexive.

Indication

From - 19999 to 32000

Refresh

5 /seg. (Filter selectable)

A/D CONVERTER

Speed

14 Readings / Second

Accuracy

16 BIT + sign (± 65.000 points)

CMRR

> 130 dB

POWER

Standard

230 Vac 50/60 Hz. Consumption 3,5W Max

Optional

115 Vac 50/60Hz. Consumption 3.8W Max

Optional

24 Vdc  (±10%) isolated. Consumption 4W Max
Isolation 1000 Vdc (Primary - Secondary)
Maximum consumption at 24 Vdc .- 265 mA
Peak current at start-up <600mA

UNIT TAS-1-VDC

Input Signal Ranges

0/650 Vdc, 0/100 Vdc, 0/10 Vdc
0/1Vdc, 0/100 mVdc, 0/20 mVdc

Note .- Although not mentioned above, the unit accepts any signal range
between 0/20mV and 0/650Vdc, including 60mV, 100mV y150mV shunt
signals. For these ranges. select the closest range and readjust with code
14 11

Accuracy

0,1% from reading ± 1 point

Thermal Drift

± 100 ppm/ºC

Input impedance

1 MOhm for Vin>1Vdc
5 MOhm

 

for Vin<1Vdc

Effective Resolution

32.000 points

UNIT TAS-1-VAC

Input Signal Ranges

0/650  Vac, 0/100 Vac

0/1 Vac, 0/100 mVac

Note .- Although not mentioned above, the unit accepts any signal range
between 0/20mVac and 0/650Vac, including 60mV, 100mV y150mV shunt
signals. For these ranges. select the closest range and readjust with code
14 11

Accuracy

0,3% of reading ± 1 point

Thermal Drift

± 150 ppm/ºC

Input impedance

1 MOhm for Vin>1Vdc
5 MOhm

 

for Vin<1Vdc

Effective Resolution

32.000 points

ENVIRONMENTAL DATA

Working Temperature

  0 ... + 50 ºC

Storage Temperature

-20 ... + 85 ºC

Humidity

  0 ... 85%, non condensaded

MECAHNICAL DATA

Dimensions

Standard 1/8 DIN
96 x 48 x 124 mm.  (3,78" x 1.89" x  4,88")

Front

Protection IP65  (NEMA 4)

Weight

0,5 Kgs

ALARMS AND RELAYS (OPTION AL2)

4 alarms on display. Alarms 1 and 2 can control Relays 1 y 2 (Option AL2)
2 Relays ON/OFF 2 A. (non inductive), 250 Vac

Note .- AL2 option not mounted on standard units

ANALÓG OUTPUT

Technical data see page 10

RS485 MODBUS OUTPUT

Technical data see page 12

PEAK / VALLEY, ETC OPTION

Technical data see page 14

UNIT TAS-1-IDC

Input Signal Ranges

0/5 Adc
0/100 mAdc

Note .- Although not mentioned above, the unit accepts any signal range
between 0/100mA and 0/5 Adc. For these ranges. select the closest range
and readjust with code 14 11

Accuracy

0,1% of reading ± 1 point

Thermal Drift

± 150 ppm/ºC

Input impedance

0.02 Ohm for I>100mA
1 Ohm for I<100mA

Effective Resolution

32.000 points

UNIT TAS-1-IAC

Input Signal Ranges

0/5 Aac
0/100 mAac

Note .- Although not mentioned above, the unit accepts any signal range
between 0/100mAac and 0/5 Aac, including X/5 and X/1 Current Transformer
signals. For these ranges. select the closest range and readjust with code
14 11

Accuracy

0,3% of reading ± 1 point

Thermal Drift

± 200 ppm/ºC

Input impedance

0.02 Ohm for I>100mA
1 Ohm for I<100mA

Effective Resolution

32.000 points

SIZES AND PANEL CUT-OUT

1234567890123456789

1234567890123456789

1234567890123456789

1234567890123456789

1234567890123456789

1234567890123456789

1234567890123456789

1234567890123456789

1234567890123456789

1234567890123456789

1234567890123456789

1234567890123456789

92mm

44mm

124mm

2mm

15mm

96mm

48mm

Summary of Contents for TAS-1-IAC

Page 1: ...CA Page 1 FEMA ELECTR NICA INTELLIGENT PANEL METERS Series TAS 1 Intelligent Panel Meters info fema es www fema es USER S MANUAL AC Current DC Current AC Voltages DC Voltages TAS 1 IAC TAS 1 IDC TAS 1...

Page 2: ...ration 6 4 6 Manual Adjustment 7 4 7 Dealing with Errors 7 4 8 Advanced Indication 7 4 9 Blocking the Keypad 7 4 10 Resetting the system Default values 7 4 11 Messages and Errors 8 5 Configuring with...

Page 3: ...Level Input Signal correction Key DP Decimal Point Position Key PROG Allows the introduction of the PROGRAMMING codes 4 digit codes as shown on page 4 These codes when validated activate different inp...

Page 4: ...t 14 14 INPUT ONERR OFF ON 1 1 Visualization 14 13 DISPLAY NULL LAST 1 1 ON OFF 1 1 OFF ON 3 2 1 FIL 2 FIL 3 FIL 1 FILT A D Q N AVERA REAL 1 1 INPUT STEP XXXXX Alarm 1 Alarm 2 Alarm 3 Alarm 4 13 11 AL...

Page 5: ...ion XXXX X 1 Press DP 2 message DP during 1 second Press 2 to fix decimal point position 2 To fix at other positions press 1 2 3 4 or 5 3 Validate changes pressinf message INPUT shows the configuratio...

Page 6: ...m activation The hysteresis points are defined on parameter HYST for each alarm Security Mode activated on alarms acting as minimum Securty Mode deactivated Output relay are inactive until the set poi...

Page 7: ...Code 14 11 Function Code 14 22 Function Resets the unit to the default manufacturing values It is needed to reconfirm the action by pressing INTRO after the message Acquisition Mode Real STEPValue 10...

Page 8: ...recovers the normal working state when situation clears Errors Errors are associated with critical situtations which disable the hability to measure the signal The error remains active on display unt...

Page 9: ...tion 32 000 points ENVIRONMENTAL DATA Working Temperature 0 50 C Storage Temperature 20 85 C Humidity 0 85 non condensaded MECAHNICALDATA Dimensions Standard 1 8 DIN 96 x 48 x 124 mm 3 78 x 1 89 x 4 8...

Page 10: ...ot need to be readjusted most of the times in case of trouble with the input signal such as loop break or sensor break the analog output signal will exhibit always the same behaviour predefined by the...

Page 11: ...essage ERROR Informs that at least one of the values is not correct and the whole configuration could not be applied Repeat the process from the begining 7 2 INSTALLING THE OPTION BOARDS The Option Bo...

Page 12: ...mposed of 1 4 bits b0 0 Value contained on register 00 00H is an indication number b0 1 Value contained on register 00 00H is not an indication Word b4 b3 b2 b1 0 Reserved 3 Ovr 6 Con 9 MErr1 12 Reser...

Page 13: ...acion MASTER and SLAVE ADDRESS 1 CHARACTER Instrument Address FUNCTION 1 CHARACTER Function 04H register read READING STARTREGISTER 2 CHARACTERS Register 00 00H Display Value NUMBER OF REGISTERS TO RE...

Page 14: ...tering the Menu Introduce the programming code 14 23 Message E INP TEK board recognized Configuring the external contacts Parameter INP_1 Contact 1 ON Active OFF NotActive Parameter FUN_T Function ass...

Page 15: ...unit to control a machine or process where injury to personnel or damage to equipment or process may occur as a result of failure of the unit External Protection Fuse to be added for 230 Vac 80mA fuse...

Page 16: ...YS COUNTERS PROCESS SIGNALS TEMPERATURES TACHOMETERS SERIAL RSREPEATERS CLOCKS CHRONOMETERS PANEL METERS AC ELECTRICAL SIGNALS DC ELECTRICAL SIGNALS TEMPERATURES LOADCELLS PROCESS SIGNALS TACHOMETERS...

Reviews: