background image

F

E

MA

 · MANUFACTURING FOR INDUSTRIAL AUTOMATION 

16

13. Configuration menu (cont.)

At the ‘

Display information

’ (

dISP

) menu select one parameter to read 

on display when the instrument is in ‘

normal mode

’ of operation. If you 

need access to more than one information, see the ‘

messages

’ function 

(see section 13.8) associated to front key ‘

LE

’ (

3

).

• 

select ‘

Measure

’ (

MEAS

) to read the value of the actual millivolts at 

signal terminals (for example: ‘

MEAS mV 7.82

’). Value is expressed in 

millivolts (see section 10.1).

 select ‘

Tare

’ (

tArE

) to read the actual value of the ‘

tare

’ parameter. 

(for example : ‘

tArE mV 1.27

’) This value is expressed in corrected 

millivolts (mV’) (see section 10.1).

 select ‘

Input signal value

’ (

InP.S

) to read the input signal value and 

the measurement units (for example: ‘

Inp mV 8.52

’). This value is 

expressed in millivolts (mV’) (see section 10.1).

 select ‘

Output signal value

’ (

out.S

) to read the output signal value 

and the measurement units (for example : ‘

Out mA 12.40

’).

 select ‘

Label

’ (

LAbL

) to read the value configured at the ‘

label

’ and 

label2

’ parameters (see section 13.9).

 select ‘

Process value

’ (

Proc

) to read the process value as scaled at 

the process parameters (see section 13.4) (for example: ‘

Proc 150.0

’).

• 

select ‘

Percentage

’ (

Prct

) to read the percentage of input signal, 

where ‘0’ is the value assigned to the ‘

input signal low

’ parameter, and 

‘100’ is the value assigned to the ‘

input signal high

’ parameter (see 

section 13.4) (for example: ‘

Prct 23.5

’).

• 

select ‘

Excitation voltage

’ (

EX.V

) to read the value of the excitation 

voltage received by the load cell. This value is read from the ‘

sense

’ 

terminals (see section 7.2) (for example : ‘

ExV 4.97

’).

• 

select ‘

Excitation current

’ (

EX.MA

) to read the value of the current 

provided through the excitation voltage terminals (for example: ‘

ExMA 14.3

’).

Display 

information

Input signal value (in mV’)

Output signal value

Label

Signal percentage

Process value

Excitation voltage (in Vdc)

Excitation current (in mA)

Tare value (in mV’)

Field

 

correction

Field correction 

low

Field correction 

low

At the ‘

Field correction

’ (

F.cor

) menu, there is access to the ‘

field 

correction

’ functions. The ‘

field  correction

’ functions allow to modify 

the ‘

input signal low

’ and ‘

input signal high

’ parameters of the ‘

Advanced 

scaling

’ menu (see section 13.4), based on the actual input signal 

measured  at  the  input.  Functions  used  to  correct  and  fine  tune  the 

slope of the load cell, by loading low and high weights and applying the 

correction low and high. Tares can are applied  after the correction.

 select the ‘

Field correction low

’ (

Fc.Lo

) function to set the actual 

input signal value at the ‘

input signal low

’ parameter of the ‘

Advanced 

scaling

’ menu. While measuring the value, the message ‘

ok

’ remains 

flashing for 5 seconds. When the measure is completed, the instrument 

returns to the ‘

Field correction low

’ (

Fc.Lo

) parameter.

 select the ‘

Field correction high

’ (

Fc.hI

) function to set the actual 

input signal value at the ‘

input signal high

’ parameter of the ‘

Advanced 

scaling

’ menu. While measuring the value, the message ‘

ok

’ remains 

flashing for 5 seconds. When the measure is completed, the instrument 

returns to the ‘

Field correction high

’ (

Fc.hI

) parameter.

‘ok’  message  flashes  while 

the  ‘field  correction’  function 

is being applied and when 

finished,  the  instrument 

returns  to  previous  menu 

entry.

13.5 Field correction

Millivolts at signal terminals (in mV)

The ‘

tare

’ value is reset to ‘0’ when a ‘

Field correction low

’ 

(

Fc.Lo

) or ‘

Field correction high

’ (

Fc.hI

) is applied.

13.6 Display information

Summary of Contents for I4 Series

Page 1: ...www fema es SERIES I4 Model I4L USER S MANUAL ISOLATED SIGNAL CONVERTER Section INDUSTRIAL 5583r00 INDUSTRIAL Weight Kg Load cell mV...

Page 2: ...ocs 5587_I4L_installation_en pdf CE declaration www fema es docs 5642_CE Declaration_I4_en pdf Warranty www fema es docs 4153_Warranty1_en pdf Web www fema es Series_I4 3 Additional information When t...

Page 3: ...Read the practical load cell information see section 7 4 Connect the input the output and the power terminals see section 9 error messages may appear in the process of connection see section 16 for ex...

Page 4: ...ead signal for changes at the excitation voltage The actual value of the excitation voltage is detected by using the sense wires Connect the sense and sense terminals 5 and 2 to the load cell to provi...

Page 5: ...g 3 load cells is the optimal way to distribute the weight on a plane although it is common to work with 4 load cells in applications with tanks hoppers and similar When working with multiple load cel...

Page 6: ...trument reads uSEr Code identifies the end of the list it follows code 199 and the list continues with code 010 Select to exit the list without applying changes To calculate the optimal input signal r...

Page 7: ...dc 7 8 9 common Vdc fuse signal sense Vexc POWER ABC 18 to 265 Vac dc isolated OUTPUT SIGNAL 789 sense common 0Vdc or passive mA current out signal 4 20mA mA current in signal 0 10Vdc or active mA cur...

Page 8: ...ion voltage terminals OUTPUT SIGNAL The output signal is configurable to 4 20mA active and passive and 0 10Vdc Table 13 Input signal ranges for load cell signals Input range Code for 4 20mA output Cod...

Page 9: ...ifications OUTPUT SIGNAL The output signal is configurable to 4 20mA active and passive and 0 10Vdc Table 14 Connection examples for millivolt signals 1 2 3 not connected mV not connected 4 5 6 not co...

Page 10: ...he input with no filter 115 mSec typ 0 to 99 with 50Hzfilter or 60Hzfilter 150mSec typ 0 to 99 with 50 and 60Hz filter 300mSec typ 0 to 99 OUTPUT SIGNAL RANGES active current output 4 20mA active max...

Page 11: ...SERIES I4 Model I4L Section INDUSTRIAL ISOLATED SIGNAL CONVERTERS www fema es 11 T h i s p a g e b l a n k...

Page 12: ...t returns to normal mode of operation HOW TO OPERATE INSIDE THE CONFIGURATION MENU Inside the configuration menu use the front keypad to move through menu entries parameters and select configuration v...

Page 13: ...ee section 13 8 The messages function ends when all the information has been displayed or front keys UP 5 or SQ are pressed The display returns to normal mode of operation 12 5 How to activate the Mes...

Page 14: ...cted a configuration code see section 13 1 the input signal has been already selected and there is no need to manually select the Mode ModE or Signal range rAnG parameters again At the Mode ModE param...

Page 15: ...ressed in x xx mA or in x xx Vdc at the Output high signal ou hI parameter configure the high output signal value This value is expressed in x xx mA or in x xx Vdc These four parameters define the rel...

Page 16: ...citation current EX MA to read the value of the current provided through the excitation voltage terminals for example ExMA 14 3 Display information Input signal value in mV Output signal value Label S...

Page 17: ...rrected millivolts mV see section 10 1 configure Input signal value InP S to on to see the input signal value and the measurement units for example Inp mV 8 52 This value is expressed in millivolts mV...

Page 18: ...and to underrange output signal to minimum value in case of input signal underrange at the On SQ on Sq parameter configure the behavior of the output signal when the operator is inside configuration...

Page 19: ...set a dead band value of 0 08mVdc All signals below 0 08mVdc will be treated as 0mVdc and the output will be 4mA the Version VEr parameter informs about the firmware version running in the instrument...

Page 20: ...oltage Excitation voltage 0 70mV 0 80mV 10mV 20mV 30mV 40mV 50mV 60mV 70mV 80mV 5mV Output range ok message flashes while the field correction function is being applied and when finished the instrumen...

Page 21: ...abel2 Alphanumerical 0 0 to 100 0 SOS mode of output On error Output to high Output to low Output standard Power filter No filter Filter for 50Hz Filter for 60Hz Filter for 50Hz and 60Hz On SQ Output...

Page 22: ...ell cELL Signal range rAnG 10 00mV Excitation voltage V EXc on Output range out 4 20mA Advanced scaling Ad Sc Tare tArE 0 00 mV Input signal low In Lo 0 00 mV Input signal high In hI 10 00 mV Output s...

Page 23: ...r motor controllers electrovalves together with signal and or control cables The use of shielded cables is recommended to prevent the coupling of environmental electromagnetic noise connected to earth...

Page 24: ......

Reviews: