Section INDUSTRIAL . ISOLATED SIGNAL CONVERTERS
13
12.
Technical specifications
INPUT SIGNAL RANGES
process
4/20 mA, 0/10 Vdc (active and passive)
excitation v15 Vdc @25 mA
thermocouples
J, K, N, E, T, R, S, C and B
conforming to ITS-90
‘Pt’ sensors
Pt100 (2 wires and 3 wires)
Pt500, Pt1000 (2 wires)
‘NTC’ sensors
resistances
ranges from 0/1 Kohm up to 0/1 MOhm
potentiometers
nominal value from 250 Ohm to 15 KOhm
ACCURACY AT 25 ºC
see for each type of signal at section 11
*
accuracy values are indicated for 4/20 mA output. For
0/10 Vdc output, add +0.05 % to indicated accuracy
values.
THERMAL DRIFT
±100 ppm/ºC (F.S.)
±0.05 ºC/ºC (thermocouple cold junction)
STEP RESPONSE
Step response associated to the configured
power filter (see section 13.10). Typical response
values to reach 99% of the output signal, as a
response to a 100% step at the signal input (see
OUTPUT SIGNAL RANGES
active current output
4/20 mA active
max. <22 mA, min. 0 mA
maximum load <400 Ohm
passive current output
4/20 mA passive
max. 30 Vdc on terminals
voltage output
0/10 Vdc,
max. <11 Vdc, min. -0.05 Vdc (typ.)
minimum load > 10 KOhm
CONFIGURATION SYSTEM
key pad + display
accessible at the front of the instrument
configuration
‘
configuration menu
’ and ‘
predefined codes
’
scalable units
scalable input ranges
scalable output ranges
scalable process display
POWER SUPPLY
voltage range
18 to 265 Vac/dc isolated
(20 to 240 Vac/dc ±10%)
AC frequency
45 to 65 Hz
consumption
<3.0 W
power wires
1 mm
2
to 2.5 mm
2
(AWG17 to AWG14)
overvoltage category
2
ISOLATION
input - output
3000 Veff (60 seconds)
power - input
3000 Veff (60 seconds)
power - output
3000 Veff (60 seconds)
ENVIRONMENTAL
IP protection
IP30
impact protection
IK06
operation temperature
from 0 to +50 ºC
storage temperature
from -20 to +70 ºC
‘
warm-up
’ time
15 minutes
humidity
0 to 95% non condensing
altitude
up to 2000 meters
MECHANICAL
size
106 x 108 x 22.5 mm
mounting
standard DIN rail (35 x 7.5 mm)
connections
plug-in screw terminal (pitch 5.08 mm)
housing material
polyamide V0
weight
<150 grams
packaging
120 x 115 x 30 mm, cardboard
Table 30
|
Response times
Signal type
No filter
50 Hz or 60 Hz filter
Both
Process
<60 mSec.
<250 mSec.
<600 mSec.
Pt100
<100 mSec.
<320 mSec.
<2 Sec.
Thermocouple
<100 mSec.
<200 mSec.
<1 Sec.
Resistances
*
<100 mSec.
<200 mSec.
<200 mSec.
*
for the 1 MOhm range, applies double the response time indicated