FEMA ELECTRÓNICA - Page 3
User’s Manual - MAG-35 Process (Model 36)
FRONT VIEW
SPAN
Light adjust of span
OFFSET
Light adjust of zero
S P A N
O F F S E T
CONNECTIONS
Power
(See label on the instrument)
EXCITATION VOLTAGE
VOLTAGE Vexc
from 10 to 24 Vdc regulated
CURRENT
50 mA max.
RIPPLE
0.2% max. (at 50/60 Hz)
READING SCALE SELECTION
Select the desired input range and the reading scale, by
selecting the jumpers on the «MM» module, as shown on table
below, and select jumpers «D» to light the decimal point (see
figure1). Potentiometer «P2» allows manual adjustment of the
display at full-scale. Potentiometers «SPAN» and «OFFSET»
at the front of the instrument are for small «on-the-field» error
corrections on the scaled display. Potentiometer «P1» is to
adjust the excitation voltage output.
INTERNAL MODULE
«MM»
P1
Vexc ADJUST
C
B
A
K
L
M
N
O
P
D
SPAN ADJUST
P2
Table2 - Span display
SPAN
JUMPER
«MM»
0 / 500
K
501 / 1000
---
1001 / 2000
L
OFFSET
JUMPER
RANGE
«MM»
0 / 500
P
501 / 1000
---
1001 / 1999
N
Table3 - Offset display
OFFSET
JUMPER
TYPE
«MM»
POSITIVE
O
NEGATIVE
M
Table1 - Signal
RANGE
JUMPER
«MM»
0/20 Vdc
---
0/200 Vdc
D
Hold
(optional)
2 1
10 9
C
B
A
- +
Common
+V
exc
Signal(Vdc)
Signed .-
D. Juncà
Position .-
Quality Manager
Barberà, 2005
DIRECTIVES
EUROPEAN DIRECTIVE FOR LOW VOLTAGE D73/23/CEE AMENDED BY D93/
68/CEE.
Equipments powered from 50 to 1000 Vac and/or from 75 to 1500 Vdc
ELECTROTECHNICAL REGULATION FOR LOW VOLTAGE (RBT) ITC21,
ITC29, ITC35.
Equipments with power supply lower than 50 Vac and/or 75Vdc.
EUROPEAN DIRECTIVE FOR ELECTROMAGNETIC COMPATIBILITY D89/336/
CEE AMENDED BY D93/68/CEE
STANDARDS
IMMUNITY
ELECTRICAL SAFETY
UNE EN 50082-1 (1997)
UNE EN 61010-1 (1996)
UNE EN 60204-1 (1997)
EMMISIONS
UNE EN 50081-1 (1993)
DECLARATION OF CONFORMITY
Manufacturer.-
FEMA ELECTRÓNICA, S.A.
Address .-
Pol. Ind. Santiga - Altimira 14 (T14 - N2)
E-08210 Barberà - BARCELONA
ESPAÑA - SPAIN
Conforming Products
Series
.-
MAG-35
Models
.-
01, 02, 03, 04, 06, 07, 08, 10, 11, 12, 15, 16, 17, 20, 21, 22,
23, 24, 25, 26, 27, 32, 35, 36, 80, 90, 95
W
e hereby declare that the above products conform to the essential protection
requirements of Directives and Standards stated below.
FIGURE1 - DECIMAL POINT SELECTION
1
.
XXX
1X
.
XX
1XX
.
X
D1 D2 D3
D1 D2 D3
MODULE
«MM»
FIGURE2 - SIZES
in mm (inches)
1234567890123456
1234567890123456
1234567890123456
1234567890123456
1234567890123456
1234567890123456
1234567890123456
1234567890123456
1234567890123456
1234567890123456
PANEL CUT-OUT
92.5 x 44.5 mm
(3.64’’ x 1.75’’)
13
(0,51’’)
102
(4,02’’)
15
(0,59’’)
5 max
(0,20’’)
48
(1,89’’)
96
(3,78’’)