getWAVELENGTHOFFSET@@
- Get the Dominant Wavelength Offset
Transmit
Receive
getwavelengthoffset@@
+/-XX
Where:
@@
represents the Fiber Number and is a number in the range 1 – 10.
+/- xx
represents the wavelength offset set of the LED in nanometers.
Description
This command is used to get the value of the Dominant Wavelength Offset set for Fiber
number
@@
.
The range of value that can be programmed is +/-99, however setting a value of greater than
+/-10 would probably indicate an error in your measurement setup
Example:
The PC transmits
getwavelengthoffset01
to the LED Analyser to instruct it to send the
programmed dominant wavelength offset for the LED positioned under Fiber No 1. The LED
Analyser will return a string +/-
xx
to the PC.
getwavelengthoffset01
+07