
FEASA LED ANALYSER
FUNCTIONAL VERSION
User Calibration Mode
getWAVELENGTHOFFSET@@
- Get the Dominant Wavelength Offset
Transmit
Receive
getwavelengthoffset@@
+/-XX
Where:
@@
represents the Fiber Number and is a number in the range 1 – 20.
+/- xx
represents the wavelength offset set of the LED in nanometers.
Description
This command is used to get the value of the Dominant Wavelength Offset set for Fiber
number
@@
.
The range of value that can be programmed is +/-99, however setting a value of greater than
+/-10 would probably indicate an error in your measurement setup
Example:
The PC transmits
getwavelengthoffset01
to the LED Analyser to instruct it to send the
programmed dominant wavelength offset for the LED positioned under Fiber No 1. The LED
Analyser will return a string +/-
xx
to the PC.
getwavelengthoffset01
+07
65
Summary of Contents for 10-F
Page 126: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Software Back to Index 126...
Page 127: ...FEASA LED ANALYSER FUNCTIONAL VERSION Test Software Back to Index 127...
Page 128: ...FEASA LED ANALYSER FUNCTIONAL VERSION User Calibration Back to Index 128...
Page 129: ...FEASA LED ANALYSER FUNCTIONAL VERSION Programming Back to Index 129...