FCS BMIC Operator'S Manual Download Page 7

 
 
 
 

 

Hand probe

 

 

For  normal  soft  ground  surfaces  the  single  pointed  extension  rod  should  be  connected,  do 
not  over  tighten  the  screw  thread.  For  direct  contact  with  underground  fittings  via  access 
covers  it  may  be  necessary  to  use  the  second  extension  rod.  To  do  assemble  this  remove 
the single rod and connect the extension rod and re-fit the pointed rod again. If the rods have 
been over tightened, spanner slots allow easy removal. 

 
 
 

Surveying

 

 

The  leak  location  can  be  narrowed  down  by  listening at accessible contact points such as 
meters,  hydrants,  valves  and  stop-taps,  which  provide  points  of  good  sound  pick-up 
particularly  if  the  pipe  is  metallic.  Use  the  hand-probe/extension  rod  to  listen  at  these 
points. 

 

If  there  are  no  accessible  contact  points  or  if  the  pipe  is  of  non-metallic  material  use  the 
microphone  foot  for  listening,  placing  the  foot  over  the  pipe  route  in  the  area  of  the 
suspected  leak.    Move  along  the  pipe  route  listening  at  each  accessible  pipe-fitting  or  at 
regular positions on the ground until you have identified the area of maximum noise level. 

 

Note

: When you are listening on pipe-fittings the location of the point of maximum noise will 

probably not indicate the leak position, only the fitting closest to the leak. The noise level will 
also appear stronger  where there is less thickness of ground or other material for it to pass 
through. The leak noise will follow the path of least resistance. 

 
 
 

Pinpointing the leak position

 

 

Pinpointing  the  leak  position  involves  a  process  of  comparing  a  number  of  leak  noises.  
Select the most suitable sensor device; the microphone foot for hard ground surfaces or the 
hand probe for soft ground. 

 

Press  the  trigger  to  operate  the  Bmic  and  adjust  the  headphone  volume  control  to  a 
comfortable  listening  level.    Once  you  have  listened  to  the  noise  through  the  headphones, 
release the trigger to switch off the amplifier and move the microphone foot or hand probe to 
the next test position. 

 

Repeat  the  sequence  to  listen  to  each  of  the  test  locations  moving  along  the  pipe  route  in 
the  direction  where  signal  strength  is  increasing.  If  the  leak  noise  level  falls  you  have 
passed  the  leak  and  should  go  back  and  reduce  the  distance  that  you  move  between 
measurements. The loudest leak noise will then indicate the location of the leak bearing in 
mind the ground conditions already mentioned. 

 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for BMIC

Page 1: ...0 BMIC Operators Guide Version 1 Warning This manual contains important safety and operating information Please read understand and follow the instructions in the manual...

Page 2: ...s Introduction Equipment List Page 3 Bmic Controls Using the Bmic 4 LED indication 4 Care of the Bmic 4 Battery warning 5 Battery replacement 5 Charging the Bmic batteries 5 Operator Guidance 7 Techni...

Page 3: ...tion and a volume sensitivity rotary control An LED indicates operation and battery status It can be used to confirm the position of underground leaks audibly from ground level sound pipe fittings pen...

Page 4: ...ipod foot on hard ground and use the probe rods on soft ground or when accessing water pipe fittings Connect the headphones to the connector and connect the microphone cable to the forward connector P...

Page 5: ...ntually need to be replaced To do this undo the four screws underneath the hand held unit and separate the two halves of the Bmic unit Gently pull out the battery holder and replace the battery When r...

Page 6: ...the water and the pipe wall as well as into the ground around the pipe The noise travels much better through hard materials so that the noise travels much further along metallic pipes than asbestos ce...

Page 7: ...When you are listening on pipe fittings the location of the point of maximum noise will probably not indicate the leak position only the fitting closest to the leak The noise level will also appear s...

Page 8: ...Ohms 4 pin Amphenol connector to control unit LED Green normal operation Red low battery power Optional accessories High specification aviation quality headphones Car cigarette lighter adapter with 12...

Page 9: ...8 Fluid Conservation Systems 1960 Old Gatesburg Road STE 150 State College PA 16803 800 531 5465 fluidconservation com MAN 080 0002 A BMIC Operators Guide Version 1...

Reviews: