background image

FLUID COMPONENTS INTERNATIONAL LLC

                             FLT93 Flow Switch Series

This page is subject to proprietary rights statement on last page

2

 

 

 

Doc. No. 06EN003312 Rev. D

Install the flow element, with the flow arrow (shown on Page 1) in the direction of media flow.  The element should be in the
center line of the process pipe or rectangular duct.  The flow arrow flat area is to be parallel ±2° with the media flow.  If the
remote control circuit option is used, the serial number of the flow element must match the serial number of the electronic
enclosure. Below are the most common instrument mounting options.

Flow Element Installation

Wiring Preparation

Before the instrument is opened to install the wiring, FCI recommends that the following ESD precautions be observed:

Use a wrist band or heel strap with a 1 megohm resistor connected to ground.  If the instrument is in a shop setting there should
be static conductive mats on the work table and floor with a 1 megohm resistor connected to ground.  Connect the instrument to
ground.  Apply antistatic agents such as Static Free made by Chemtronics (or equivalent) to hand tools to be used on the
instrument.  Keep high static producing items away from the instrument such as non-ESD approved plastic, tape and packing
foam.
The above precautions are minimum requirements to be used.  The complete use of ESD precautions can be found in
the U.S. Department of Defense Handbook 263.
Unscrew the instrument’s lid.  Remove the control circuit by prying up on the metal transformer tabs (under the serial
number label) while rocking the transformer back and forth.  This exposes the wiring block shown on the next page.

Flanged Integral Instrument Shown in a Customer Process

NPT Integral Instrument Shown in a Customer Process

FLT93L Local Inline Mount

Be sure the instrument is configured correctly for the applied input power.  If Factory Mutual
specifications were ordered, the input power should be 115 VAC and the instrument jumpers should be set
for 115 VAC as shown on Page 4.  Otherwise, the power input has been set for 220 VAC unless otherwise
specified.

Caution

Summary of Contents for FLT93 Series

Page 1: ...lement Serial Number Also Showing FlowArrow Located near the FE enclosure It is also on the enclosure tag Electronics Serial Number It is also on the transmitter s FT enclosure tag C Recommended insta...

Page 2: ...1 megohm resistor connected to ground Connect the instrument to ground Apply antistatic agents such as Static Free made by Chemtronics or equivalent to hand tools to be used on the instrument Keep hi...

Page 3: ...Remote Wiring Pictorial FLT93 Integral Wiring Diagram Maximum Distance for AWG Connection 10 ft 50 ft 100 ft 250 ft 500 ft 1000 ft 3m 15m 31m 76m 152m 305m AC Power 22 22 22 20 18 16 Relay 6A 22 16 12...

Page 4: ...OR FLT S AND J14 IS STANDARD FOR FLT93 F 0 75 WATTS 0 21 WATTS OFF OFF 0 49 WATTS 0 25 WATTS JUMPER ALARM NO 1 RELAY DE ENERGIZED WITH LOW FLOW LOW LEVEL DRY OR HIGH TEMPERATURE RELAY DE ENERGIZED WIT...

Page 5: ...h alarm can be setup for one of three applications flow level interface or temperature The following application specific adjustment procedures are generic and can be used for setting either or both a...

Page 6: ...turn the set point adjustment potentiometer clockwise until the LED turns on With the LED on slowly turn the potentiometer counterclockwise one turn past the point at which the LED just turns off The...

Page 7: ...kwise until the LED turns off With the LED off slowly turn the potentiometer clockwise one half turn past the point at which the LED just turns on The potentiometer may have up to one quarter turn of...

Page 8: ...No 2 counterclockwise until the LED turns off and then slowly clockwise just until the LED turns on 8 If this is the only flow application alarm to be setup then skip to the Continue With the Air Gas...

Page 9: ...ontinue With the Air Gas Flow Procedure 1 Slide the mode switch to the RUN position 2 Establish the normal process flow condition For low flow alarm setups the status LED should be off For high flow a...

Page 10: ...etecting Wet Condition high level alarm If the status LED is on turn the set point adjustment potentiometer R26 for alarm No 1 or R25 for alarm No 2 slowly counterclockwise just until the LED turns of...

Page 11: ...ge is out of tolerance NOT OK The serial numbers on the control board and the sensing element must match If the serial numbers are OK then contact FCI Field Service c The Voltage is between 1 and 6 vo...

Page 12: ...secrets and proprietary information which is the property of Fluid Components International LLC FCI Disclosure of this data to you is expressly conditioned upon your assent that its use is limited to...

Reviews: