10
4.
USING THE SYSTEM
Setting up the UFU (to be done prior to starting
the procedure)
1.
Turn on the UFU by pressing down on the ON-OFF control
switch located on its top-left edge for approximately 5 seconds,
until the boot screen appears.
If only the battery-charge ICON
appears, the battery is
fully depleted and the UFU must be recharged prior to use.
NOTE: A full charge for a fully-discharged UFU takes about three
hours.
Unless the UFU is fully-depleted, the Android
‘Home’ screen
will appear after 20 seconds.
After another 10 seconds, the
TruNode®
application will launch
and the UFU will be ready to detect the
TruNode®
probe.
Caution: If the battery charge is below 50%, turn off the
UFU and charge it prior to use. Do not use the Charger in
the operating room.
1.
The UFU will announce a greeting followed by its charge
status. Verify the UFU is at least 50% charged prior to starting
the procedure.
2.
When an image of the probe keypad with a flashing
‘
M
’
button is displayed, the UFU is ready to link with a
TruNode®
Probe.
3.
Clamp the UFU securely to an IV pole in a location that is easy
to see by the user.