system manual Part I : installation & operation
4.3
WB25 model
WR25 model
HR25 model
TXtal
(
µ
s )
23.60
29.80
32.66
Table 4.1: typical values of
TXtal
(a) Button
(b) Contextual menu
Figure 4.11:
Repetition Frequency
adjustment helpers (WR25 Dazzler
TM
typical values)
4.2.3
Trigger to Laser
adjustment procedure
As shown in
and
, the precise synchronization of the acoustic with the
optical wave requires that the
Trigger to Laser
value be adjusted with the same accuracy. The
following steps explain how to proceed.
1. Set
Trigger to Laser
to the value corresponding to the setup. If necessary, measure it
with a photodiode and an oscilloscope.
2. If working in
Trig on Previous
mode, adjust
Repetition Frequency
to the value corre-
sponding to the setup.
3. Send the diffracted beam to a spectrometer (
REMINDER : amplifiers should not be used
until the
Trigger to Laser
value is correctly adjusted.
).
4. Set the phase option to polynomial and the ”order 3” and ”order 4” to zero. Use the
”delay” dial to center the signal on the time graph.
5. Increase the second order phase coefficient (”order 2”) until the signal covers the full
window on the software time graph (
6. Measure the diffracted spectrum.
7. Increase the
Trigger to Laser
value until the measured spectrum is truncated. Then,
decrease the value until the measured spectrum is truncated again.
8. Set the
Trigger to Laser
value to the mean of these two values.
9. Re-check that the diffracted spectrum is not truncated.
V3.00 - 8