![FASTLITE DAZZLER Manual Download Page 32](http://html1.mh-extra.com/html/fastlite/dazzler/dazzler_manual_3588631032.webp)
DAZZLER
TM
system manual Part I : installation & operation
4.2
Figure 4.2: Left: acoustic wave position in the crystal when the optical wave propagates through
the crystal (acoustic wave unclipped), Right: Equivalent spectrum and time graphs
4.1.3
Effects of incorrect trigger settings
When
Trigger to Laser
is greater than the optimum value (
), or more exactly when
Trigger Delay
is set too short
, the optical pulse occurs when the trailing part of the acoustic
wave is present in the crystal: the wave was launched too early, hence its leading part is missing.
Figure 4.3: Incorrect
trigger
settings:
Trigger Delay
too short
When
Trigger Delay
is set too long, the acoustic wave is not yet entirely in the crystal when
the light pulse arrives. The trailing part of the wave is missing.
4.2
The
Trig&Mode
panel
The Dazzler
TM
software provides a user-friendly interface to enter the delays
T
p
,
TXtal
,
Trigger
to Laser
,
Trigger Delay
correctly based on experiment parameters.
Before opening the
Trig&Mode
panel, check that the single mode operation is selected on the
software front panel (one RF cycle is started by one external trigger, see
The
Trig&Mode
panel opens by clicking on menu
“setup
→
trig and mode”
.
5
Trigger to Laser
depends of the laser construction and is a “given” system constant while
Trigger Delay
can be adjusted
V3.00 - 8
th
April 2019