DazScope
manual
Chapter 1 : Introduction
1.2
What the DazScope is not
Although the DazScope was designed and developed to be as user-friendly as possible, it
cannot
fully replace an intelligent operator. More precisely, the DazScope cannot be expected to correct
spectral phases which exceed the pulse shaping capabilities of the Dazzler used in conjunction with
the DazScope. This is why the optimization procedure is semi-automatic and requires, in the basic
mode, a few operations from the user.
The DazScope is based on a characterization technique described in
. It must be
underlined that as any phase-only method, this algorithm is
not
a pulse measurement technique
comparable with the FROG or SPIDER techniques: the DazScope is only relevant for pulses
with sufficiently regular spectra and sufficiently smooth spectral phases. The pulses delivered by
chirped pulse amplification (CPA) laser systems typically verify such specifications. However, a
double pulse structure, for example, will not be correctly measured nor corrected.
FASTLITE-9 september 2008