FasTest FasCal Series Operating Instructions Download Page 2

 

OPERATING INSTRUCTIONS

 

 

    Roseville, MN 

     Ph: 1-800-444-2373 
     Fax: 651-645-7390 

     www.fastestinc.com 

                                                                                                                                                                                                                                                                                   WP101 Rev C  9/4/2012             

TEST PIECE 

MEDIA PORT 

SQUEEZED HANDLE 
POSITION 

REMOVAL 

INSERTION 

PILOT PORT 

THUMB VALVE 

PILOT PORT 

OPERATION: 

1. Squeeze the handle and insert the gripping collets of the 

FasCal

 into the 

female test port threads until the main seal o-ring contacts the component 
to be tested. 

 
 
 
 
 
 

2. Release  the  handle  of  the 

FasCal

,  expanding  the  collets  to  grip  the  test 

port threads. 

 Threaded collets lock into female threaded port, drawing in the sealing o-

ring creating a leak tight seal for pressure testing, evacuation, charging or 
filling operations. 

 
 
 
 
 
 
 
 
 
 
 

3. Introduce test media through media port for the duration of test.  

See chart below for media port, fitting, torque specifications. Thread seal-
ant must be used..  
Note: The torque limit is typically achieved 2-3 turns past finger tight of the 
fitting. 

Media Port Size (NPT/BSPP) 

Media Port Torque Limit (lbf-ft) 

1/8 

12 

1/4 

21 

3/8 

28 

1/2 

35 

3/4 

46 

90 

 
 
 

4. When  test  is  complete  relieve  test  pressure  and  squeeze  the  handle  to 

disengage threaded grip collets from test piece. 

 
 
 
 
 

5. The  previous  four  steps  of  connecting  also  apply  to  the  operation  of  the 

Pneumatic and Thumb Valve operated connectors. 
 

6. Both the Pneumatic and Thumb Valve connector require 60-120 psi. Both 

operators have 1/8” NPT pilot ports. 
 

7.  

Pneumatic  Valve  Connector:

  This  requires  external  control  provided  by 

the user to engage and disengage the collets pneumatically. 

 
 
 
 
 

8.  The  Thumb  Valve  connector  controls  collet  engagement  by  pressing  the 

Thumb Valve on connector. 

 

 

WARNING: 

 

FasTest

  Connectors  are  designed  ONLY  for  the  threads  meeting  the 

specifications  listed  in  the 

FasTest

  literature.  DO  NOT  use  in  other 

threads.

 

 

Connectors  are  NOT  designed  for  permanent  connections  and  are  for 
temporary connections only.

 

 

DO NOT EXCEED pressure rating as marked on connector.

 

 

Verify  proper  connection  BEFORE  pressurizing.  Use  only  in  a  safe 
environment.  Safety  chains  and  guards  are  recommended  for  all 
applications.

 

 

DO  NOT  move  connector  while  under  pressure.  Damage  or  injury  may 
result.

 