4.  To return to normal weight display (gross weight), 

remove any items from the weighing platform and 
press the

 

T

 key again to re-zero the scale.

VII.  Calibration

Calibration and adjustment are performed at the factory. 

However, the weighing range can shift slightly during 

shipping. Also, due to changes in local acceleration of gravity 

around the world, adjustment is recommended when moving 

the scale from one location to another. To check calibration:

1.  Place the required calibration weight on the scale 

and note the displayed value.

2.  If the displayed value is not within acceptable 

tolerance (see chart below), adjustment is required.

Model

Acceptable 

Tolerance (±)

Calibration 

Weight

Recommended 

Accuracy Class

MS-600

0.2g

500g

OIML M2

VIII.  Adjustment

1.  Turn the scale OFF.

2.  Press and hold the 

M

 key then tap the 

T

 key to turn 

the scale on.

3.  Release the 

M

 key once the display shows a ran-

dom number (ISN).

4.  Tap the 

T

 key again. The display will show “Zero” 

followed by the required calibration weight.

5.  Place the required weight on the center of the 

platform and press the 

T

 once more.

6.  The display will show “PASS” and return to normal 