4. To return to normal weight display (gross weight),
remove any items from the weighing platform and
press the
T
key again to re-zero the scale.
VII. Calibration
Calibration and adjustment are performed at the factory.
However, the weighing range can shift slightly during
shipping. Also, due to changes in local acceleration of gravity
around the world, adjustment is recommended when moving
the scale from one location to another. To check calibration:
1. Place the required calibration weight on the scale
and note the displayed value.
2. If the displayed value is not within acceptable
tolerance (see chart below), adjustment is required.
Model
Acceptable
Tolerance (±)
Calibration
Weight
Recommended
Accuracy Class
MS-600
0.2g
500g
OIML M2
VIII. Adjustment
1. Turn the scale OFF.
2. Press and hold the
M
key then tap the
T
key to turn
the scale on.
3. Release the
M
key once the display shows a ran-
dom number (ISN).
4. Tap the
T
key again. The display will show “Zero”
followed by the required calibration weight.
5. Place the required weight on the center of the
platform and press the
T
once more.
6. The display will show “PASS” and return to normal