background image

CALIBRATION 

When and How to Calibrate the Scale 

Calibration  may be required when it is initially installed, if the scale is  moved to a 

substantial distance. 

This is necessary because the weight of a mass in one location is not necessarily the 

same  in  another  location.  Also,  with  time  and  use,  mechanical  deviations  may 

occur. 

How to calibrate 

 

 

Press and  hold  the 

[MODE]

 key,  then  press and  hold 

[ON/OFF] 

key.  The  display 

will then show 

[CAL]

 then show five digits.   

 

Press

 [MODE] 

and the display will then flash 

[500g]

.   

 

Place  the  calibrating  weight  on to the  pan.  Wait  for  the  stable  indicator  to  display, 

then press 

[MODE]. 

The  display will show ‘-----‘ and then return to showing five 

numbers. Calibration is complete.     

 

Press 

[ON/OFF]

  to  turn  the  power  off,  Press 

[ON/OFF  ]

  to  turn  the  power  on   

again,  Place  the  full  weight 

(500g)

  on  the  pan  to  check  calibration.  If  it  is  not 

correct, repeat the calibration procedure.

 

 
 
 
 
 

PREPARATION 

Installing Batteries/Connecting the AC/DC Adaptor                   

Batteries 

  Battery                                                                                Remove the battery compartment   
  compartment                                                                      cover and insert four batteries(R6P/LR6 

                                                                                              /AAA) into the battery compartment, 

                                                                                              taking extreme care that the polarities   

                                                                                              plus and minus is observed. 

                                                                                         

 

 
                                                             

    4 x LR6/AAA size 

                                                                            Dry batteries. 
  Battery compartment cover                               

 

AC/DC 300mA Power Adapter (optional) 

                                                                       

          Plug the AC/DC Adapter to the 

      AC/DC Adaptor                                                           

 

adapter socket on the side. 

      socket                                                                     

       

The AC input requirement could 

                                                                                              be 110,120,220,230 or 240Volts 

                                                                                              (50/60Hz) depending on the area 

                                                                                              where used, so please verify that 

                                                                                              the adaptor is correct 

                                                                               

 

2.Setting up the scale/balance 

 

!

 

CAUTION 

Avoid installing the scale/balance in direct sunlight that may cause discoloration or 

malfunctions.  Place  your  scale/balance  on  a  firm  weighing  table  so  that  the 

scale/balance is level. (The scale/balance will not perform accurately when it is not 

level.) 

 

Place the scale/balance on the firm surface that is flat and level for an accurate 
weighing. 

 

Before weighing 

Whenever  possible,  please allow the balance to warm up for 10  minutes after  first 

Reviews: