Model 88C Instruction Manual
Do not use solvents to clean the cell.
6.
Using the nomograph (Figure 4-1), convert the resistivity in ohm-meters to
concentration of chlorides.
4.2
Semisolid Samples
This procedure explains how to measure resistivity of semisolid samples, such as
filter cakes and mud solids.
1.
Prepare samples of uniform moisture content.
2.
Fill the slot on the outer surface of the cell completely with a semisolid sample.
3.
Follow steps 2 through 6 from 4.1 Fluid Samples.