Setup Enabled Features
The DFM may have one or more optional features enabled. These features are only enabled at the
factory when the customer purchases the feature.
Note:
Because of LED dimming, press the SET button twice to advance to each feature. Pressing the
Set button once brightens the LEDs.
Press On-Off
•
Place the DFM in standby mode by pressing the ON-OFF button if the power to
the DFM is not activated. The LEDs are visible in standby mode.
•
Press the SET button to advance to the first enabled feature.
Note: If a feature is not enabled, it will not appear in the display.
Automatic Shut-Off on Time Feature
OFFT:
Y or N appears on the display. You have the option to temporarily
disable this feature by pressing the UNITS or RESET button to change
the Y to an N.
The time setting screen displays hhh:mm (hours: minutes). Any time
value can be set from 0:01 to 168:00 hr:min.
1.
Set the
minutes
value first. Press the UNITS button to increase the
minutes value or the RESET button to decrease the minutes value.
2.
Press the SET button to move to the hours set-up screen. Press the
UNITS button to increase the
hours
value or the RESET button to
decrease the hours value.
NOTE:
If the hours value has changed pressing the SET button
allows the customer to go back to the minutes setup screen.
3.
Press the SET button to move to the next feature.
Automatic Shut-Off on Volume Feature
OFFV:
Y or N appears on the display. You have the option to temporarily
disable this feature by pressing the UNITS or RESET button to change
the Y to an N.
Any volume value can be set from 0.01E00 to 9.99E99
Note:
The “V” in the “OFFV” is actually a “U.” This is because of the
segments of the LED display.
NOTE:
The automatic shut-off on volume feature generally is not
enabled if the automatic shut-off on time has been enabled. However, it
is physically possible to have both enabled. The shut-off feature that is
most restrictive will occur first and thus terminate the sampling event.
1.
Set the
exponent
value first. Press the UNITS button to increase the
exponent value or the RESET button to decrease the exponent value.
2.
Press the SET button to move to the
digits
set-up screen. Press the
UNITS button to increase the digits value or the RESET button to
decrease the digits value.
NOTE:
If digits to the right of the decimal point have changed,
pressing SET takes the customer back to the exponent setup screen.