Manual LactoFAN

2

 

11 of 30

 

 

Hold breath 
While the countdown, hold the breath. 

 
If the patient is able no more to hold the breath before the countdown is over, press the OK-button to 
continue.

 

 

Blow into the device 
Put the mouthpiece to the lips tightly and exhale slowly and as long as 
possible (20 s). 

 
The system got a big flow resistance and supports the subject in achieving a long expiration time. 
It is important that the patient exhales completely, because the relevant part of breath for determining the H

2

 

concentration comes from the tip of the lung. 
Press the OK-button to continue. 

 

Wait for maximum reading 
There may be a delay of some seconds before any reading appears. Do 
not turn off the unit during this time. 

 

Because of there is a certain delay in the chemical sensor’s output, the reading is monitored for further 

increasing. 
After a decay of 2ppm but not later than after 30 seconds the device switches to the next screen. 

 

Confirm measurement 
Press the OK button to store the reading. 

 
If you press the cancel button this measurement is discarded. You can try again after the recovery time is 
over. 

 

Take out the mouthpiece 
Pull the filter mouthpiece out of the device. Instruct the subject to keep 
the mouthpiece for the following measurements of the breath test. 

 

 

 

Was the mouthpiece detection disabled in the settings, the OK-button is shown. 

 

 

Press it when finished. 

 

Summary of Contents for LactoFAN2

Page 1: ...Instructions for Use H2 Breath Test Device LactoFAN2 V 1 16 1 21...

Page 2: ...ect data and the consequences arising thereof In case of doubt the German version shall apply Copyright Copyright 2010 2014 Fischer ANalysen Instrumente GmbH All rights reserved Duplication modificati...

Page 3: ...ironment 20 Spare parts consumables and service 20 H2 breath test protocols with LactoFAN2 23 Specifications 24 Manufacturer declaration of electromagnetic compatibility of H2 breath test device Lacto...

Page 4: ...ported then lactose intolerance may be diagnosed The LactoFAN is based on an electrochemical fuel cell which works through the reaction of hydrogen with an electrolyte at one electrode and oxygen from...

Page 5: ...No LF2 AC 01 xxx 1 pcs mouthpiece adapter with one way valve translucent plastic Order No HZ ADV 1 pcs one way cardboard Mouthpiece Order No HV MU150 Additional included in delivery mini USB cord sof...

Page 6: ...t is feasible but after that the device have to be stored in a well aired room again Do not touch the device with hands just disinfected 1 display with touch screen 2 flange with mouthpiece detection...

Page 7: ...own low change main battery soon 2 sensor back up battery charge indicator shown low device service needed 3 remaining days until next calibration press to start calibration 4 start measurement 5 chan...

Page 8: ...ion If there are any warnings displayed e g battery low please follow the instructions in the regarding chapter of this manual Check whether the device is in a calibrated state see page 7 If not perfo...

Page 9: ...of 30 Explain to the subject the right usage of the mouthpiece Correct The mouthpiece fits closely to the subject s lips no breath escapes Wrong The rim indents the subject s lips the respiratory air...

Page 10: ...o set the table into which the measurement data of the current patient will be stored This button will abort the measurement always Data table Verify to have selected the right patient table then pres...

Page 11: ...Wait for maximum reading There may be a delay of some seconds before any reading appears Do not turn off the unit during this time Because of there is a certain delay in the chemical sensor s output t...

Page 12: ...account not only the next measurement time according to your breath test protocol but the time needed to call the patient too Press the cancel button if you finished the measurements for the current p...

Page 13: ...ailable So you can save print and process your data easily Functional tests In daily operation it is not necessary to submit the LactoFAN2 any functional tests This does not affect the requirements fo...

Page 14: ...knob anti clockwise until the ball in the flow indicator is between the two marks This will then supply a gas flow of approximately 0 3 l min Now slowly turn the Apply this flow for 50 seconds and the...

Page 15: ...Manual LactoFAN2 15 of 30 100 ppm hydrogen in nitrogen Flow indicator Control valve Reducing connector for calibration Plastic tubing...

Page 16: ...ilability of the device Please take the batteries out of the device in case you intend not to use it for a longer period of time You have to set the internal clock when the device is returned to opera...

Page 17: ...measurement range of 300 ppm H2 is not possible to be acquired Cause Same as for 001 016 calibration range exceeded The sensors sensitivity alteration in respect to the last calibration is out of the...

Page 18: ...tery exhausted 256 sensor battery empty Send the device to the FAN service workshop see page 20 device temperature too high or low 032 temperature too low 064 temperature too high Wait a sufficient ti...

Page 19: ...ant in your country concerning dealing with and disposal of the batteries you use to run the device Keep the device out of reach of the patient while it is connected to a PC The LactoFAN should only b...

Page 20: ...in the range from 470 MHz to 800 MHz by cellular phones and by electromagnetic interferences see declaration of conformity Spare parts consumables and service Orders for spare parts and consumables or...

Page 21: ...N2 21 of 30 1 LactoFAN2 LF2 2 battery or accumulator set LF2 AE 04 LF2 AE 01 3 one way filter mouthpiece LF2 AC 01 xxx 4 CD with software and manuals H soft 5 USB cable LF2 AE 03 6 carry case LF2 AM 0...

Page 22: ...22 of 30 Manual LactoFAN2 7 calibration gas 100ppm H2 HV Lacto CG 8 regulator for calibration gas with reducing connector HZ CV HZ RedC 9 reducing connector for calibration HZ RedC 7 8 9...

Page 23: ...tion 0 5 12 5 30 180 HOB 15 ppm H2 malabsorption Glucose Bacterial Overgrowth 1 50 20 120 HOB 15 ppm H2 overgrowth Lactulose Bacterial Overgrowth 0 5 10 20 120 HOB 20 ppm H2 overgrowth Lactulose Intes...

Page 24: ...etreibV or into another equivalent directory Range 0 to 300 ppm vol vol Resolution 1 ppm Display 2 8 color AMOLED panel with capacitive glass touch field Shell Polycarbonate ABS elastomeric coating Di...

Page 25: ...uding living quarters and such ones which are connected directly to a public power supply that supplies also buildings used for living purposes Table 2 Guidelines and manufacturer declaration Electrom...

Page 26: ...e applicable for all cases The propagation of electromagnetic values is influenced by absorptions and reflections of buildings objects and people a The field strength of stationary transmitters such a...

Page 27: ...ion in the corresponding column P is the maximum nominal power of the transmitter in watt W according to the specifications of the transmitter manufacturer NOTE 1 For 470 MHz and 800MHz the higher ran...

Page 28: ...28 of 30 Manual LactoFAN2 CE declaration...

Page 29: ...f the FAN GmbH is a Medical Device Class Im in the scope of the valid European laws regarding medical devices The Act on Medical Devices Medical Devices Act Medizinproduktegesetz MPG of 7th August 200...

Page 30: ...n 28 Declaration of Quality 29 Electromagnetic Compatibility 25 disinfectant agents 3 environmental conditions 6 manufacturer 20 mouthpiece 19 mouthpiece how to use 9 nose clip 10 operating temperatur...

Reviews: