The WMA-01 and 01LF amplifiers contain 
an  over-voltage  protection  for  daily  use  in 
the  laboratory  and  are  long-term  short-
circuit  proof.  The  input  (indicated  “Input” 
on  the  label),  output  (indicated  “Output”) 
and DC power supply (Indicated “24V DC 
180mA”)  connections  are  all  standard 
BNC. 
 
The  amplification  is  20x  (fixed).  The 
amplifiers have an output voltage range of 
–125V  to  +125V.  The  24V  DC  power 
supply  requirement  offers  the  possibility 
for battery powered operation.  

 

 

Safety 

 
-  This  product  is  able  to  produce  over 
125V,  but  the  50kΩ  output  impedance 
limits the current to a rather safe maximum 
of  2.5mA.  This  voltage  is  painful,  but  will 
not  be  lethal  (under  normal  circum-
stances). Nevertheless take care. This is a 
high  voltage  unit  and  safety  measures 
should be taken accordingly. 
-  The  internal  circuitry  of  the  amplifier 
operates  at  high  voltage.  Only  qualified 
personnel  from  Falco  Systems  should 
service it. 
- When the amplifier is turned on or off, a 
short  voltage  spike  may  appear  at  the 
output which may damage circuitry already 
connected to it. 
-  Replace  the  fuse  with  250V  315mAT 
5x20mm rated fuses only. 
-  This  product  is  only  suitable  for  indoor 
use  in  a  class  II  environment  (domestic, 
light industrial).   
-  The  power  supply  connected  to  the 
WMA-01/01LF should be double insulated 
(class  II)  or  a  SELV  type  (SELV  =  Safety 
Extra Low Voltage, typically a battery). 
- The WMA-01/01LF is a class III product. 
Do not connect  the  amplifier  to  the  mains 
safety  earth.  If  this  is  done,  the  system 
becomes a class I apparatus with class III 
isolation,  which  means  that  the  protective 
insulation  may  be  inadequate.  Doing  so 

(only  under  laboratory  conditions)  is  fully 
at the users’ risk. Risk of electric shock! 
-  The  WMA-01/01LF  should  not  be  used 
above 2000m altitude to prevent electrical 
breakdown. 
- This product should only be cleaned with 
a  soft,  slightly  moist  cloth.  Unplug  the 
WMA-01/01LF  from  the  DC  power  before 
cleaning. 
 
 

Input protection

 

 
A  low-noise  amplifier  like  the  WMA-01  or 
WMA-01LF  can  never  be  made  fully 
insensitive to input overload conditions, as 
this  would  limit  the  performance  of  the 
amplifier to an unacceptably low level. The 
absolute  maximum  input  voltage  is  ±15V. 
Warning:  exceeding  this  spec  may  cause 
permanent malfunction of the amplifier and 
is not covered by the warranty! 
For  normal  operation,  input  voltages 
should  remain  in  the  -6.25V  to  +6.25V 
range,  resulting  with  an  amplification  of 
20x  in  an  output  voltage  swing  of  more 
than -125V to +125V.  

 

Ouput protection

 

 
The  WMA-01/01LF  has  been  designed  to 
be  fully  stable  with  all  capacitive  loads.  It 
has  been  optimized  for  its  step-response, 
but  is  also  a  very  good  linear  and  sine-
wave  amplifier.  No  significant  overshoot 
occurs at any capacitive load. 
Overloading 

or 

short-circuiting 

this 

amplifier will not break down the amplifier, 
due  to  the  50kΩ  current  limiting  resistor 
that has been employed at the output. 
 
Although the amplifier cannot be damaged 
by  a  short-circuit  condition  or  capacitive 
loading, two situations should be avoided: 
- Connecting a charged capacitor. 
- Using high inductance values (coils). 
  
 

 
 
 
 

 

Summary of Contents for WMA-01

Page 1: ...e g for use in a Faraday cage Its internal high voltage generator is inductor free and hence does not spread magnetic interference in sensitive experimental setups Its large voltage range makes it an...

Page 2: ...mains safety earth If this is done the system becomes a class I apparatus with class III isolation which means that the protective insulation may be inadequate Doing so only under laboratory condition...

Page 3: ...connection can cause overshoot in the oscilloscope reading The amplifier is capable of amplifying above 9kHz and should not be used for telecommunication as described in the R TTE directive 95 5 EC T...

Page 4: ...t voltage Figure 3 Frequency response at 200mVpp output voltage Frequency Hz Amplitude Vpp 250 200 150 100 50 0 10 0 10 1 10 2 10 3 10 4 10 5 WMA 01LF WMA 01 Amplitude V pp Frequency Hz WMA 01 WMA 01L...

Page 5: ...5 Figure 4 250Vpp 1kHz square wave Figure 5 Full scale 10kHz square wave Figure 6 30mVpp 1kHz small square wave output signal...

Page 6: ...0pF 470pF 2 2n 100pF 1nF 10nF Figure 7 Full power 1kHz square wave response with different capacitive loading conditions The 50k output resistor limits the speed at which the capacitor can be charged...

Page 7: ...Noise less than 10mVpp or equivalently less than 3mVrms and typical DC offset 1 2mV for the WMA 01 The high voltage generator frequency around 150kHz contributes considerable to the overall noise Thi...

Page 8: ...th typical for the WMA 01 1 2mV offset typical Noise 0 1mVrms noise over the full bandwidth typical for the WMA 01LF 1 5mV offset typical Input impedance 100k Output impedance 50k Stability stable wit...

Page 9: ...function of the product resulting from use or operation in other ways than specified in the user manual Malfunctioning due to misuse or abuse of the product Malfunctioning occurring after changes or r...

Reviews: