Falco Systems WMA-01 User Manual Download Page 1

 

 

 

USER MANUAL 

 

 
 

 

 
 
 

● High voltage: 20x amplification up to +125V and -125V output voltage  

 

● DC to 20kHz large signal bandwidth and 50kΩ output resistance 

 

● Power supply 24V DC, battery powered operation possible 

 

● No overshoot with capacitive loads: bandwidth changes automatically 
to ensure stability 

 

● Short-circuit protected output 
 

 

 
About this manual 

 
This user manual is an integral part of the 
Falco  Systems  WMA-01  and  WMA-01LF 
high  voltage  amplifier  products.  Please 
read  it  carefully  and  pay  attention  to  the 
recommendations and instructions for safe 
use. 

 

 

The WMA-01 high voltage 
 amplifier: general description 

 

The  Falco  Systems  WMA-01  and  WMA-
01LF  are  high  voltage  amplifiers  that  run 
from an 24V DC power supply and can be 
battery powered if required, e.g. for use in 
a  Faraday  cage.  Its  internal  high  voltage 
generator is inductor-free and hence does 
not  spread  magnetic  interference  in 
sensitive  experimental  setups.  Its  large 

voltage range makes it an excellent choice 
for  use  with  MEMS  devices,  EO-
modulators,  piezo  actuator  positioning 
systems,  beam  steering  components,  and 
many more.  
The  amplifiers’  full-power  frequency 
response  has  a  -3dB  corner  frequency  at 
20kHz. The input impedance is 100kΩ and 
the output 50kΩ. 
 
The  output  noise  of  the  WMA-01  is 
<3mVrms.  For  extremely  low  noise 
applications  where  bandwidth  is  not  an 
issue, the dedicated WMA-01LF version is 
available. This amplifier contains an output 
low pass filter capacitor which reduces the 
bandwidth  to  50Hz,  but  makes  the  noise 
drop  to  0.1mVrms.  This  bandwidth-limited 
amplifier  is  ideal  for  MEMS  or  piezo 
positioning 

with 

manual 

voltage 

adjustment requiring extremely low noise. 

Falco Systems WMA-01           High Voltage Amplifier DC – 20kHz 

Summary of Contents for WMA-01

Page 1: ...e g for use in a Faraday cage Its internal high voltage generator is inductor free and hence does not spread magnetic interference in sensitive experimental setups Its large voltage range makes it an...

Page 2: ...mains safety earth If this is done the system becomes a class I apparatus with class III isolation which means that the protective insulation may be inadequate Doing so only under laboratory condition...

Page 3: ...connection can cause overshoot in the oscilloscope reading The amplifier is capable of amplifying above 9kHz and should not be used for telecommunication as described in the R TTE directive 95 5 EC T...

Page 4: ...t voltage Figure 3 Frequency response at 200mVpp output voltage Frequency Hz Amplitude Vpp 250 200 150 100 50 0 10 0 10 1 10 2 10 3 10 4 10 5 WMA 01LF WMA 01 Amplitude V pp Frequency Hz WMA 01 WMA 01L...

Page 5: ...5 Figure 4 250Vpp 1kHz square wave Figure 5 Full scale 10kHz square wave Figure 6 30mVpp 1kHz small square wave output signal...

Page 6: ...0pF 470pF 2 2n 100pF 1nF 10nF Figure 7 Full power 1kHz square wave response with different capacitive loading conditions The 50k output resistor limits the speed at which the capacitor can be charged...

Page 7: ...Noise less than 10mVpp or equivalently less than 3mVrms and typical DC offset 1 2mV for the WMA 01 The high voltage generator frequency around 150kHz contributes considerable to the overall noise Thi...

Page 8: ...th typical for the WMA 01 1 2mV offset typical Noise 0 1mVrms noise over the full bandwidth typical for the WMA 01LF 1 5mV offset typical Input impedance 100k Output impedance 50k Stability stable wit...

Page 9: ...function of the product resulting from use or operation in other ways than specified in the user manual Malfunctioning due to misuse or abuse of the product Malfunctioning occurring after changes or r...