FagronLab™ UVGI-80  – User Manual  |  6

Error

Troubleshooting

Work indicator is not lit

Check the connection of power

Check the fuse condition

UV light is not working

Check Ballast working condition 2A (Ф5×20)

UVC lamp is damaged

Fan is not working

Check the fan motor

Check the fan power connection

10. Trouble Shooting

11. Maintenance

12. Warranty

  UV life-time is 5000hours. Please replace the UV lamp in time according to your personal disinfection plan

  To store the device, unplug from power and cover with a clean cloth 

This device is under warranty and free from defects in materials and workmanship, under normal use and service, for a 
period of 12 months from the date of invoice (excluding consumable accessories). The warranty is extended only to the 
original purchaser. Warranty is not valid on device which has been damaged on account of improper installation, improper 
connections, misuse, accident or abnormal conditions of operation. If the warranty has been expired, Fagron would still 
responsible for repair with relative charges.

For claims under the warranty please contact your local supplier.

FagronLab™ UVGI-80

  Please unplug the device from the power and contact your local supplier, when the following occurs: 

  

A  

When the power cord or plug is damaged or worn

  

B

  If liquid is spilled into the unit

  

C

  If the sterilizer is being exposed to rain or water

  

D

  If the sterilizer is broken or the frame is damaged

  

E

  If the sterilizer changes significantly in performance

  The service life of the whole device is five years. If used for more than five years, the aging of components and power 

cords may cause personal safety accidents. A complete Service must be done after this period. 

  Ultraviolet germicidal lamps should be used to detect the irradiance according to the time of use, and the irradiance of 

the radiation is <70uW/cm2 (power ≥30W). The UV germicidal lamp should be replaced regularly (lifetime more than 
5000 hours)

Summary of Contents for UVGI-80

Page 1: ...FagronLab UVGI 80 User Manual www fagronlab com ...

Page 2: ......

Page 3: ...4 3 Bactericidal Factors and Strength 4 4 Microbial Category 4 5 Technical Parameter 4 6 Functions and Use 5 7 Installation 5 8 Operation 5 9 Attention 5 10 Trouble Shooting 6 11 Maintenance 6 12 Warranty 6 13 Service 7 14 Transportation and Storage 7 15 Circuit diagram 7 ...

Page 4: ...ricidal factor used ultraviolet irradiation UV C through short wavelength ultraviolet UV lamps meet the technical requirements achieving 254nm of strength Germicidal lamp is a special type of lamp which produces ultraviolet UVC light This short wave ultraviolet light disrupts DNA base pairing causing formation of pyrimidine dimers and leads to the inactivation of bacteria viruses and protozoa Mode...

Page 5: ...e room should be closed during disinfection Power must be turned off during installation and maintenance It is strictly forbidden to cover or block the air inlet and air outlet of the device Do not insert anything into the device It may cause fire or electric shock Do not spill any liquid on the sterilizer The power should be turned off before cleaning the device During maintenance or cleaning tur...

Page 6: ...n improper connections misuse accident or abnormal conditions of operation If the warranty has been expired Fagron would still responsible for repair with relative charges For claims under the warranty please contact your local supplier FagronLab UVGI 80 Please unplug the device from the power and contact your local supplier when the following occurs A When the power cord or plug is damaged or wor...

Page 7: ...tomer care representative with the following information Description of problem Serial Number of the device Methods and procedures adopted to resolve the problems Your contact information The transportation should avoid rain mechanical collision and mixing with corrosive gas substances The machine should be stored in a dry well ventilated non corrosive gas warehouse 15 Circuit diagram ...

Page 8: ...Together we create the future of personalized medicine Fagron Hellas 12 km N R Trikala Larisa P C 42100 P O Box 32 Trikala Greece T 30 24310 83633 5 F 30 24310 83615 www fagron gr U VA SU M 0 1 ...

Reviews: