12
APPENDIX E
POSSIBLE ERRORS ON POWER UP
The following error messages can appear on the CRT of the VNP400 on power up.
RAM ERROR
It occurs when detecting a RAM memory fault.
Turn the DRO off and back on.
*
If the error persists, Call the service department of FAGOR AUTOMATION.
*
If the error disappears, check the machine parameters and part programs since the error
may have altered their contents.
GENERAL PARAMETER ERROR
It occurs when detecting a malfunction at the general machine parameters.
The DRO will initialize this table with the factory settings. See the section on “Reset
Parameters”.
Re-enter the appropriate values.
PARAMETER ERROR, AXIS 1
PARAMETER ERROR, AXIS 2
PARAMETER ERROR, AXIS 3
PARAMETER ERROR, AXIS 4
It occurs when detecting a malfunction at the machine parameters for the indicated axis.
The DRO will initialize this table with the factory settings. See the section on “Reset
Parameters”.
Re-enter the appropriate values.
VARIABLE VALUE ERROR
It occurs when detecting an improper value in any internal variable of the DRO.
Check the machine parameters and part-programs since the error may have altered their
contents.
PROGRAM ERROR
It occurs when detecting an internal memory fault and when the DRO finds non-numeric
characters in the program number or when the indicator for total number of blocks of a
program has been modified.
Check all programs.