Factair F8104 Operating Manual Download Page 6

 

Home Screen Display Icons 

 

 

 

 

 

 

 

 

 

 

 

Standard Equipment List 

  Optional Extras 

F8104 Safe-Air Tester 

TUB0053 – 8103560 Dräger Oil Impactor – Box of 10 

USB Download Cable 

TUB0003 – 6728371 Dräger Oil Tubes – Box of 10 

Stylus Pen 

F3002 High-Pressure Regulator 

Wall Mounting Plate 

F1946 Connection ¼ BSP inlet adaptor 

ASSY0013 supply hose with rectus 21 
plug and socket connections 

F2193 Safe-Air Tester to Dräger/Scott compatible plug  

F1946 Connection ¼ BSP inlet adaptor 
with rectus 21 socket 

F2194 Connection adaptor terminating in a Rectus Plug 

 

F2195 Connection adaptor terminating in a Instantair 
plug 
 

Selected Test Standard 

Settings 

Battery Charge 
& Charging 
Indicator 

Installer         
Access Status 

SIM Card 
Signal 

Red – No 
Signal 

Green - 
Signal 

 

Ethernet 
Connectivity 
status 

SD Card 
Memory 
Status 

Unlocked 

Locked 

Supply 
Pressure 

By pressing 
the icon you 
can change 
display units 
between 
barg and  
psig 

Supply 
Temperature 

By pressing 
the icon you 
can change 
display units 
between °C 
and °F 

Time and 
Date 

By pressing 
the icon you 
can adjust 
the date and 
time 

Sensor 
Status 

 = OK 

---------- 

 = Sensor 

Out of 
Calibration 

 

only : 

Dewpoint close 
to 
measurement 
limit

 

---------- 

 

= Sensor 

Fault 

 

only : 

Last test failed 
or not yet 
tested

 

Summary of Contents for F8104

Page 1: ...0 F8104 SAFE AIR MONITOR OPERATING MANUAL FAC QAM 221 Issue B 04 20...

Page 2: ...1...

Page 3: ...the Instrument 4 1 Selecting a test standard 4 2 Configuring the main alarm 4 3 Configuring the pre alarm 4 4 Auxiliary inputs outputs 4 5 Sensors 10 10 11 14 14 15 5 Exporting the Data Log 16 Append...

Page 4: ...n monoxide and oxygen infrared LED sensor for the carbon dioxide and a Photo Ionisation Detector PID for Volatile Organic Compounds The moisture is measured with an electronic dewpoint meter and the i...

Page 5: ...0Hz Mains Input PC USB Connection Dr ger Impactor Oil Mist Tube Test Port On Off Button Compressed Air Inlet Ethernet Connection 2 no 4 20mA Input 2 no 4 20mA Output 2 no Alarm Volt Free Contacts Righ...

Page 6: ...ection adaptor terminating in a Rectus Plug F2195 Connection adaptor terminating in a Instantair plug Selected Test Standard Settings Battery Charge Charging Indicator Installer Access Status SIM Card...

Page 7: ...le Organic Compounds VOC Nitrogen Oxide NO Sulphur Dioxide SO 0 40 PPM 0 10 PPM 0 10 PPM 10 6 eV Photoionisation Detector Electrochemical NO and NO2 sensors Electrochemcial Moisture 100 C to 20 C PDP...

Page 8: ...unit is free from vibration during use The F8104 Safe Air Monitor requires a 230V 1ph 50Hz mains power supply The maximum compressed air supply pressure for the F8104 Safe Air Monitor is 10 barg If th...

Page 9: ...ed The duration of the commissioning delay can be later adjusted in the installer menu If the sensor readings have already stabilised you can commence immediate sensor monitoring by pressing the Commi...

Page 10: ...is clears contaminants from previous tests and ensures that a representative sample is being tested Once the purge has finished the Dr ger Impactor oil tube should be inserted into the test port and t...

Page 11: ...standard Enter user log in password installer The standards menu will then display the available test standards As well as medical standards the F8104 can also test breathing air to EN12021 2014 This...

Page 12: ...to discard press the waste basket icon 4 2 Configuring the main alarm To enter the set up mode press this icon which is located in the top right hand corner of the screen near the battery icon Enter...

Page 13: ...e Sounder an audible tone will be emitted in the event of any of the readings exceeding their minimum and maximum value This is an optional factory fitted feature which allows a beacon to be fitted on...

Page 14: ...fitted in the instrument the SMS Message feature can be enabled by selecting this tick box When activated a message will be sent when any of the readings reach their alarm values The SMS alarm message...

Page 15: ...ted an SMS message 4 4 Auxiliary Inputs and Outputs The F8104 can display readings from external sensors which have a 2 wire 15V 4 20 mA output The description units of measure and the 4 and 20mA para...

Page 16: ...15 4 5 Sensors The order in which sensors are displayed on the home screen can be set you can also choose which sensors you want to be displayed Any changes need to be saved to take effect...

Page 17: ...4 connect the USB cable between the F8104 and your PC and open the Safe Air Tester Results F8000 series software With the F8104 successfully connected the status light will turn green The unit will th...

Page 18: ...press the button to proceed If you can determine with a high degree of certainty the specific compressor lubricant being used then refer to the test times chart provided on Factair s website marked fo...

Page 19: ...roceed as described in section 4 and at the end of it you will be prompted to enter the result from the tube The result will then be displayed on the final test result screen and stored in the instrum...

Page 20: ...s circumference this allows the user to easily and quickly identify minute quantities of oil For tests that pass with an oil concentration of less than 0 05mg m3 the screen will remain blank For tests...

Page 21: ...e to break the tip 3 Repeat steps 1 and 2 with the other end of the tube Using the tube tip cutter with the Dr ger oil tube 1 Using the Tube Tip Cutter place the ampoule section in the platform at the...

Page 22: ...ed their expiry date they should be disposed of using one of the following methods Used Tubes Submerge the tube s in a beaker or metal container filled with water and allow to soak for 24 hours Treat...

Page 23: ...er fitted in a breathing air system Suggested Test Limits for VOCs In EN12021 2014 it states compressed gas for breathing shall not contain contaminants at a concentration which can cause toxic or har...

Page 24: ...ior to fitting detector tubes clean if required Always remove the ends of the detector tubes cleanly using the correct tube tip cutter Do not use pincers or other devices Always remove both ends of th...

Page 25: ...in poor quality air being supplied for an extended period Failure of the compressed air aftercooling will result in air entering the filtration at too high a temperature this will cause the filtration...

Page 26: ...ation No discoloration Odour None None Note There is also a requirement to conduct a visual particulate test and this should be free from visible particles in a 75 L sample 175 L for surgical air Abov...

Page 27: ...26 Water mg m3 to PPM v v Conversion Charts...

Page 28: ...27 Factair Ltd 49 Boss Hall Road Ipswich Suffolk IP1 5BN UK Tel Sales 44 0 1473 746400 Tel Hire 44 0 1473 746444 Email enquiries factair co uk www factair co uk...

Reviews: