Warning!
For ATEX/IECEx GROUP II applications, ONLY Duracell Plus MN1500,
Duracell ID1500, Energizer E91 or Panasonic LR6XWA cells may be used
with the iCAM502 (see Table 1).
Warning!
For ATEX/IECEx GROUP I applications, ONLY Duracell Plus MN1500 and
Duracell ID 1500cells may be used with the iCAM502 (see Table 1).
Warning!
For ATEX/IECEx GROUP II ONLY, the user shall not carry more than a
maximum of 2 spares of the approved cells listed above into the
hazardous area. Precautions shall be taken to prevent short circuiting of
un-protected cell terminals. Discharged cells shall be removed from the
hazardous area.
Warning!
For ATEX/IECEx GROUP I applications, the changing of cells in the
hazardous area or the transportation of spare cells into the hazardous
area is NOT PERMITTED.
Warning!
For ATEX/IECEx applications, NOTE THE DIFFERENT TEMPERATURE
CLASSSIFICATION of T3 at 60
o
C when Energizer E91 or Panasonic
LR6XWA cells are used (see Table 1).
Warning!
The user shall NOT mix different cells in ANY application
Warning!
For ATEX/IECEx GROUP II applications, note that the rated ambient
temperature range of the equipment is from -20
o
C to 60
o
C.
Warning!
For ATEX/IECEx GROUP I applications, note that the rated ambient
temperature range of the equipment is from 0
o
C to 60
o
C.
Warning!
Any connection to the iCAM502 USB port must NOT be made in a
hazardous area.
Warning!
If any damage to the iCAM502 is found which may invalidate the required
safety minimum IP rating (IP54 for group I applications), (IP20 for group
II), the iCAM502 shall be removed from the hazardous area.
Important
Any repairs or replacement of parts MUST be performed by the
manufacturer or its nominated sub contractor or agent.
Important
The iCAM502 contains no user serviceable parts. The main part of the
enclosure must not be opened under any circumstances.