Extech Instruments SDL550 User Manual Download Page 4

4

   

                                          

SDL550-EU-EN-V1.2-10/11

 

Selecting Meter Functions 

The SDL550 can measure Moisture Content (%RH), Dew Point Temperature (

°

C/

°

F), Wet Bulb 

Temperature (

°

C/

°

F), and Thermocouple Temperature (

°

C/

°

F). Press and hold the FUNCTION button 

for 1.5 seconds to begin scrolling through the functions. Release the FUNCTION button when the 
desired function is displayed. Refer to the table below. 

Functions 

Primary display 

Secondary display 

Unit of measure 

Moisture Content 

Relative Humidity  

rh  briefly displayed then the 
Moisture Probe Temperature 

RH% and 

o

C or 

o

Dew Point 

Temperature* 

dp display icon 

°

C or 

°

Wet Bulb 

Temperature* 

_et display icon 

°

C or 

°

Air Temperature 

Thermocouple Temperature 

tp briefly displayed 

°

C or 

°

Measurements 

Moisture Content Measurements 

1.  Connect the Moisture probe to the jack at the top of the meter on the right 

2.  Switch the meter ON by pressing and holding the power button for 1.5 seconds 

3.  Press and hold the FUNCTION button to select the Moisture Content mode. When the Moisture 

Content mode is selected, the display icon rh is briefly shown 

4.  Carefully remove the protective cap from the moisture probe. The probe tip is extremely sharp, 

use caution 

5.  Insert the probe tip into the material under test far enough to cover the sensor’s openings 

6.  The meter will now display moisture content in %RH on the primary (upper) LCD digits 

7.  The secondary (lower) LCD digits will show the temperature measured by the moisture probe’s 

built-in temperature sensor. Use the Setup mode to change the temperature units of measure 

8.  Note that the probe uses a built-in filter. When the filter needs replacing (becomes soiled or 

defective), the measurement readings require a longer stabilization time. Please contact Extech 
for replacement filters 

Dew Point and Wet Bulb Temperature Measurements 

1.  Connect the Moisture probe to the jack at the top of the meter on the right 

2.  Switch the meter ON by pressing and holding the power button for 1.5 seconds 

3.  Press and hold the FUNCTION button to select the Wet Bulb or the Dew Point mode. When the 

Dew Point mode is selected, the display icon dp is shown on the secondary display area. When 
the Wet Bulb mode is selected, the display icon _et is shown on the secondary display area. 

4.  Carefully remove the protective cap from the moisture probe. The probe tip is extremely sharp, 

use caution 

5.  Insert the probe tip into the material under test far enough to cover the sensor’s openings 

6.  The meter will now display the Dew Point Temperature or Wet Bulb Temperature on the primary 

(upper) LCD digits. Use the Setup mode to change the temperature units of measure 

Summary of Contents for SDL550

Page 1: ...User s Guide Moisture Content Meter Plus Dew Point Wet Bulb and Thermocouple Temperature functions SD Card real time datalogger Model SDL550 RH Moisture Content Meter SDL550 ...

Page 2: ...national Safety Symbols This symbol adjacent to another symbol or terminal indicates the user must refer to the manual for further information Meter Description 1 Thermocouple Temperature Probe input jack 2 Moisture Sensor input jack 3 Primary LCD Display digits 4 Unit of Measure for Primary Display 5 HOLD and Backlight key 6 MAX MIN key 7 SET and Clock key 8 PC interface jack 9 Reset button 10 Po...

Page 3: ...ng the backlight ON or OFF unless the beeper is disabled Moisture Content Sensor The supplied Moisture Content sensor connects to the meter via the DIN jack at the top right of the meter Note that the utility end of the sensor is located beneath the protective cap use great caution when removing the cap and when handling the sensor as the sensor head is extremely sharp The sensor uses an internal ...

Page 4: ...terial under test far enough to cover the sensor s openings 6 The meter will now display moisture content in RH on the primary upper LCD digits 7 The secondary lower LCD digits will show the temperature measured by the moisture probe s built in temperature sensor Use the Setup mode to change the temperature units of measure 8 Note that the probe uses a built in filter When the filter needs replaci...

Page 5: ...asure Data Hold To freeze a displayed reading on the LCD momentarily press the HOLD button the HOLD icon will appear above the reading To exit the Data Hold mode press the HOLD button again Max Min Reading Record For a given measurement session this meter can record the highest MAX and the lowest MIN readings for later recall 1 Press the MAX MIN button momentarily to access this mode of operation ...

Page 6: ...omatically switches out of the Setup mode if no key is pressed within 7 seconds 5 The available Setup parameters are listed below Additional detailed information is provided below this list dAtE Set the clock Year Month Date Hours Minutes Seconds SP t Set the datalogger sampling rate 0 to 3600 Seconds PoFF Automatic power off management Enable or disable the auto power off function bEEP Set the be...

Page 7: ...umerical Format comma or decimal European and USA numerical formats differ The meter defaults to USA mode where a decimal point is used to separate units from tenths i e 20 00 The European format uses a comma i e 20 00 to separate units from tenths To change this setting 1 Access the dEC parameter 2 Use the arrow buttons to select USA or EUro 3 Press ENTER to confirm setting 4 Press and hold the S...

Page 8: ... recommended that the card be formatted and the logger s clock set to allow for accurate date time stamping during datalogging sessions Refer to the Setup Mode section for SD card formatting and time date setting instructions European and USA numerical formats differ The data on the SD card can be formatted for either format The meter defaults to USA mode where a decimal point is used to separate ...

Page 9: ...Up to 99 spreadsheet documents each with 30 000 readings can be stored in this folder 7 When datalogging begins a new spreadsheet document named HTB01001 xls is created on the SD card in the HTB01 folder The data recorded will be placed in the HTB01001 xls document until 30 000 readings are reached 8 If the measurement session exceeds 30 000 readings a new document will be created HTB01002 xls whe...

Page 10: ...ing and holding the POWER key for at least 1 5 seconds If using the power adaptor unplug the adaptor and then plug it back in again to power the meter Battery Replacement and Disposal When the low battery icon appears on the LCD the batteries must be replaced Several hours of accurate readings are still possible in this condition however batteries should be replaced as soon as possible 1 Remove th...

Page 11: ...Display update rate Approx 1 second Data Output RS 232 USB PC computer interface Operating Temperature 0 to 50 C 32 to 122 F Operating Humidity 85 R H max Auto Power OFF After 10 minutes of inactivity can be disabled Power Supply Six 6 1 5 VDC batteries optional 9V AC adaptor Power Consumption Normal operation backlight datalogger OFF approx 3 5mAdc With backlight OFF and datalogging ON approx 28m...

Page 12: ...nds up to 1300 C 2372 F however thermocouple sensor ranges vary greatly be sure to select a thermocouple rated for the expected temperature measurement ranges Accuracy specified for meter only Thermocouple probes add additional measurement error The above specifications are tested under an environmental RF Field Strength lower than 3 V M and a frequency lower than 30 MHz Copyright 2011 Extech Inst...

Reviews: