SD800-EU-EN- V2.4   1/13 

Advanced Settings 

The SET function is used to accomplish the following: 

 

Format the SD memory card  

 

Set the date and time 

 

Set the sampling time 

 

Set the beeper sound ON/OFF 

 

Set the SD card Decimal character 

 

Select the Temperature units 

 

Set the RS232 data output ON/OFF 

 

Set CO

2

 height compensation in meters 

 

Set CO

2

 height compensation in feet 

1.  Press and Hold the SET button for >2 seconds to enter the setting mode. The first function (Sd 

F) will appear in the display. Press the SET button to step through the seven functions. Use the 

 and 

 buttons to adjust the selected function. Use the “LOGGER” button to step through 

fields within a function. In the SET mode, if no button is pressed within 5 seconds the logger 

will revert to the standard mode. 

2.  Sd F – Format the SD card. Press the 

 button to select yES or no. For yES, press the Enter 

button to format the card and erase all existing data. 

3.  dAtE – Set the date and time. Press the 

 or 

 buttons to adjust the selected field. Press the 

Enter button to store the value and to step through the various fields. 

4.  SP-t – Set the sample rate. Press the 

 button to select the desired sample rate and press 

Enter to store the selection. The selections are: 5, 10, 30, 60, 120, 300, 600 seconds and 

AUTO. In AUTO, the data will be stored every time there is a temperature change of >1 

degree, > 1 %RH or > 50 ppm. 

5.  bEEP - Set the beeper ON or OFF. Press the 

 button to select ON or OFF and press Enter to 

store the selection. 

6.  dEC - Set the SD card Decimal character. Press the 

 button to select USA (decimal) or Euro 

(comma) and press Enter to store the selection. 

7.  t-CF - Set the Temp. unit to °F or °C and press ENTER to store the selection. 
8.  rS232 - Set the RS232 data output ON/OFF. Press the 

 button to select ON  or off and press 

Enter to store the selection 

9.  HIgh- – Set the height compensation for CO

2

 in meters. Press the 

 or 

 buttons to adjust the 

selected field to the height (altitude) of the measurement site. This adjustment will improve 

measurement accuracy. 

10.  HIghF - Set the height compensation for CO

2

 in feet. Press the 

 or 

 buttons to adjust the 

selected field to the height (altitude) of the measurement site. This adjustment will improve 

measurement accuracy. 

11.  ESC – Exit the setting mode. Press the SET button to return to normal operation. 

System RESET 

If a condition appears where the CPU does not respond to keystrokes or the meter seems frozen, 

press the RESET button on the side of the datalogger (use a paper clip or similar pointed object) to 

return the meter to a working state. 

RS232 Interface 

A serial output is provided to link the meter to a PC’s serial port. Call Technical Assistance for more 

information on this interface. 

Summary of Contents for SD800

Page 1: ...User s Guide CO2 Humidity Temperature Datalogger Model SD800 ...

Page 2: ...Carbon Dioxide Temperature and Relative Humidity Datalogger date time stamps and stores readings on an SD card in Excel format for easy transfer to a PC Selectable data sampling rate 5 10 30 60 120 300 600 seconds Product Description 1 Humidity Temperature Sensor 2 Humidity Display 3 Temperature Display 4 CO2 Display 5 LOG ENTER button 6 SET button 7 AC adaptor socket 8 CO2 probe socket 9 TIME but...

Page 3: ...advanced features section of this manual to set the clock Note The default data structure uses a decimal point as the numeric decimal indicator See the advanced features section of this manual to change this to a comma Note If the SD memory card is not installed EMPTY will appear in the display Note Displayed error messages The memory card is full or there is a problem with the card The battery is...

Page 4: ...0 000 3 After 30 000 columns a new file CHA01002 XLS is created This is repeated every 30 000 columns until CHA01099 XLS At this point a new folder CHA02 is created and the process is repeated CHA10 is the final folder Transferring Data to a PC 1 Remove the memory card from the datalogger and plug it into the SD card slot on the PC 2 Launch Excel and open the data file on the memory card from with...

Page 5: ... seconds and AUTO In AUTO the data will be stored every time there is a temperature change of 1 degree 1 RH or 50 ppm 5 bEEP Set the beeper ON or OFF Press the button to select ON or OFF and press Enter to store the selection 6 dEC Set the SD card Decimal character Press the button to select USA decimal or Euro comma and press Enter to store the selection 7 t CF Set the Temp unit to F or C and pre...

Page 6: ...off the cover 2 Replace the six AAA batteries use alkaline heavy duty type observing polarity 3 Replace and secure the cover 4 You as the end user are legally bound EU Battery ordinance to return all used batteries disposal in the household garbage is prohibited You can hand over your used batteries accumulators at collection points in your community or wherever batteries accumulators are sold Dis...

Page 7: ...P s b l 3 5 0 g 0 4 2 r e t e M t h g i e W Dimension Meter 132 x 80 x 32mm 5 2 x 3 1 x 1 3 Probe 132 x 38 x 32mm 5 2 x 1 5 x 1 3 Range Resolution Accuracy 0 0 to 50 0 C 0 1 C 0 8 C Temperature 32 0 to 122 0 F 0 1 F 1 8 F H R 4 0 7 o t 0 1 Relative Humidity 70 to 90 0 1 4 of reading 1 RH m p p 0 4 m p p 0 0 0 1 1000 to g n i d a e r f o 5 m p p 0 0 0 3 CO2 3000ppm 1ppm 250ppm typical Note Above sp...

Reviews: