Extech Instruments RH520A User Manual Download Page 7

                                                                   7                          

RH520A-en-GB_v4.8   6/19

 

 

Getting Started 

POWER 

1.

 

The RH520A runs on battery power or AC adaptor (5VDC 500mA). Battery power consists of 
three (3) ‘AA’ batteries. See the Battery Replacement section of this manual when 
changing/installing the batteries. Note: Batteries and adaptor are supplied. 

2.

 

Plug the AC adaptor in the AC adaptor jack shown in the Description section. The batteries 
will act as battery back-up in the event of an AC power failure. 

3.

 

Once the batteries are installed or the adaptor properly connected, the RH520A will begin 
displaying. 

4.

 

In the unlikely event that the AC power AND the battery back-up fails, the RH520A will 
require a “Programming Reset” to reset the display. Press the RESET button located inside 
the battery compartment.  “Reset” clears all settings. Time, Date and the logging interval 
need to be entered.  The measurement data and alarm history, however, will still be retained 
in the non-volatile memory. 

5.

 

The 5-segment battery life indicator allows the user to track the status of the battery. When 
all 5-segments are dark, the battery is fully charged. Segments switch off as the battery ages. 
When the battery indicator has only one segment left the batteries must be replaced 
immediately. 

NOTE:  Always set the Date, Time, and Sample rate immediately after batteries have been 

installed or after the RESET button in the battery compartment has been pressed.  
Then press the Data Reset button on the rear face of the meter for >4 seconds to 
clear any recorded data. 

MOUNTING THE RH520A 

The RH520A can be used in the following ways: 

1.

 

Placed on a table top where the swivel keypad is employed as the table stand 

2.

 

Wall mounted using the rear mounting holes 

MEASUREMENT PROBE 

The probe is attached to the meter with a 1-meter cable and can be used stored in the RH520A 
probe cradle or remotely. The probe cable length can be easily extended with an RJ45 cable and 
a gender changer. 

PUSH-BUTTON LOCK-OUT SECURITY FEATURE 

The RH520A can be secured from tampering by having its push-buttons locked. 

1.

 

Press for approx. 1 second and release the UPDOWNLEFT, and SET arrow buttons 
simultaneously to lock-out the push-buttons. 
Note: The security lock-out can only be performed from the main view screen. If a scroll key 
or the time button was pressed prior to setting up the lock-out, the user must scroll back to 
the main screen. 

2.

 

Press for approx. 1 second and release the UPDOWNLEFT, and SET arrow buttons 
simultaneously to restore the operation of the push-buttons 

Summary of Contents for RH520A

Page 1: ...USER MANUAL Paperless Humidity Temperature Chart Recorder Model RH520A Additional User Manual Translations available at www extech com...

Page 2: ...n Use protective gloves when necessary Do not short circuit batteries or discard them in a fire FCC Compliance This equipment has been tested and found to comply with the limits for a Class B digital...

Page 3: ...odule jack 7 AC adaptor jack 8 DATA RESET button clears recorded measurement data and alarm history 9 Temperature units select switch C or F 10 PROGRAMMING RESET button clears programming changes but...

Page 4: ...ge meter 3 Vertical axis scale limits 4 Cursor Alarm indicators 5 Push button lock out status indicator 6 Time display 7 Relative Humidity RH measurement graph 8 Date display 9 Function indicators 10...

Page 5: ...n with other buttons to set new parameter values Stores new parameter values and returns to the Standard View TEMP Used in combination with other buttons to set the vertical graph range Used in combin...

Page 6: ...xit TEMP RH View Dew Point temp VIEW to exit ALARM Step back through ALARM points ALARM selects next Alarm VIEW to exit SET TEMP ALARM Enters Set TEMP Alarm mode ALARM selects HIGH LOW SET to save SET...

Page 7: ...ator has only one segment left the batteries must be replaced immediately NOTE Always set the Date Time and Sample rate immediately after batteries have been installed or after the RESET button in the...

Page 8: ...RE o C o F The oC oF switch located inside the battery compartment is used to select the unit of measure for temperature displays SETTING THE VERTICAL RESOLUTION FOR THE TEMPERATURE GRAPHIC DISPLAY 1...

Page 9: ...displays will switch off 5 The projected TIME and DAYS representing the moment the internal memory will be full based on the sample rate is displayed in the TIME and DATE display areas 6 Use the UP D...

Page 10: ...H alarm limit Use the LEFT RIGHT arrow buttons to step through the decades 4 Press the ALARM button The MIN low alarm indicator will switch on 5 Use the UP DOWN arrow buttons to increment decrement th...

Page 11: ...Time Date and numerical measurement display fields Each press of the RIGHT ARROW button will move the cursor to the right When the cursor reaches the right side limit the graph will move to the left H...

Page 12: ...the MAX or MIN indicator is ON under the RH reading digits the display is showing the MAX or MIN relative humidity ALARM DISPLAYS 1 Press the ALARM button from the Standard View to analyze the most re...

Page 13: ...time of day 3 The temperature and humidity readings for the selected Time will display 4 Press VIEW to return to the Standard View DEW POINT DISPLAY Press the TEMP and the RH buttons simultaneously t...

Page 14: ...AYS that the internal memory will be full based on the current sample rate is displayed in the TIME and DATE display areas 4 To change the sample rate refer to the programming section of this manual 5...

Page 15: ...r allows the user to track the status of the battery When all 5 segments are dark the battery is fully charged Segments switch off as the battery ages When the battery indicator has only one segment l...

Page 16: ...imum 3 4mA 5 VDC Minimum Output Voltage 2 5 VDC Power supply Three 3 AA batteries and AC adaptor included DC 5V 500mA 3 5mm x 1 35 mm plug Battery life 4 weeks typical using alkaline batteries Operati...