Introduction 

Congratulations  on  your  purchase  of  the  Extech  PH220  waterproof  pH/mV  meter.  This 
microprocessor-based device with tactile push-buttons is battery operated and ideal for field use. 
The PH220 simultaneously displays pH and temperature. The rubber holster which protects the 
meter has a magnetic backing for convenient mounting. The built-in memory (25 readings) allows 
STORE, RECALL, and CLEAR functionality.  Careful use of this meter will provide years of reliable 
service. 
 

Specifications 

 

Range (Resolution)

 

Accuracy

 

Temperature 

 

32 to 194

o

F (0.1°<100

o

; 1

 o

 >100

o

0 to 90

o

C  (0.1°)

 

± 1.8

o

F  

± 1.0

o

C

 

pH

 

0.00 to 14.00pH (0.01pH)

 

± 0.02pH

 

-1999mV  to  1999mV 

± 2mV 

Note 1: Accuracy does not include pH buffer temperature 

coefficient 

Note 2: Accuracy specified from 0.02pH to 13.98pH 

Temperature Compensation

 

32 to 194

o

F (0 to 90

o

C) Automatic ‘ATC’ 

Power

 

9 volt alkaline battery 

Display

 

Dual LCD display for simultaneous pH and temperature 
readings 

Over-range or open indication

 

Displays "HHH" for over-range or open input conditions 
Displays “LLL” for under-range 

pH Calibration points

 

4.00, 7.00, 10.00 (1, 2, or 3 point calibration can be 
performed) 

Auto Power OFF 

10 minutes from last key-press (can be disabled) 

Low Battery Indication

 

Fill bars in the battery display icon 

 represent 

battery power status. The fewer the bars, the weaker the 
battery power. 

Dimensions / Weight

 

4.25 x 3.00 x 1.18" (108 x 75 x 30mm) / 4.8 oz. (135g) 

Accessories supplied with kit

 

Battery, suction cup holder, clip holder, stick electrode for 
model PH220-S, standard electrode with 39” cable for 
model PH220-C. 

Summary of Contents for PH220

Page 1: ...USER MANUAL Waterproof Palm pH Meter Model PH220...

Page 2: ...y does not include pH buffer temperature coefficient Note 2 Accuracy specified from 0 02pH to 13 98pH Temperature Compensation 32 to 194oF 0 to 90oC Automatic ATC Power 9 volt alkaline battery Display...

Page 3: ...perature Note that there are other more specialized displays discussed later in the appropriate sections of this manual Order of Operations 1 Calibrate the instrument as described in the Calibration s...

Page 4: ...ibrated at least once each day pH Calibration Procedure 1 Place the electrode into a 7 00 pH buffer solution Press and hold the CAL button until CAL appears in the display 2 The instrument automatical...

Page 5: ...also display when the electrode is disconnected 8 When the test is complete rinse the electrode with distilled or deionized water Automatic Temperature Compensation ATC The millivolt output of a pH el...

Page 6: ...the recall mode is accessed 3 While in the RECALL mode use the STORE RECALL button to scroll through the stored readings The most recent readings will show first LAST IN FIRST OUT 4 Press and hold th...

Page 7: ...the top of the meter ensure that the electrode is disconnected 3 Pry the battery compartment up from the top of the meter using a screwdriver or coin at one of the two indentations see diagram 4 Repl...

Page 8: ...60oC for 30 minutes Strong Alkaline measurement Soak in 0 1 M HCL overnight Deteriorated Gel layer Replace electrode Protein coating on electrode surface Soak in 1 gm Pepsin dissolved in 100ml of 0 1...

Page 9: ...bration for most of our products Contact us for information on calibration and repair availability refer to the contact information below Annual calibrations should be performed to verify meter perfor...

Reviews: