17 MG325-en-GB_v1.1 3/16
10.
Specifications
Function Range Resolution
Accuracy
DC Voltage
±600V
0.01V to 1V
(2.0% r 3 digits)
AC Voltage
0~600V
0.01V to 1V
(1.5% r 5 digits)
Specified from 5% of range to 100% of range
Input impedance: 10M
Frequency 45~400Hz (unspecified for frequency >400Hz)
Input protection: 600V DC and AC rms
Frequency
45~1kHz 0.1Hz
(0.1% r 3 digits); 45~450Hz
Frequency measurements are shown for ACV tests in the auxiliary display area
Measurements 450Hz~1kHz are for reference only; accuracy not specified
Continuity
0.01~100
(@ 20mA test
current)
0.01~0.1
(1.5% r 5 digits)
0.01~100
(@ 200mA test
current)
0.01~0.1
(1.5% r 4 digits)
Continuity threshold 30
(programmable ON/OFF continuity buzzer)
Open Circuit Voltage: 5V approximately
Resistance
0.001k~10M
0.001k
.~ 0.1M
(3.0% r 3 digits)
Capacitance
0.1nF~500
µ
F
0.1nF~0.1
µ
F
(5.0% r 5 digits)
Insulation Resistance Measurements
Test Voltage
Range
Min. Resolution
Accuracy
50V
0.00M~0.99G
1.00G~10.0G
0.01M
~ 0.01G
0.01G
~ 0.1G
(3% + 3 digits)
(3% + 3 digits) reading;
4%/G
100V 0.00M~0.99G
1.00G~20.0G
0.01M
~ 0.01G
0.01G
~ 0.1G
(3% + 3 digits)
(3% + 3 digits) reading;
2%/G
250V 0.00M~0.99G
1.00G~50G
0.01M
~ 0.01G
0.01G
~ 0.1G
(3% + 3 digits)
(3% + 3 digits) reading;
0.8%/G
500V 0.00M~0.99G
1.00G~100G
0.01M
~ 0.01G
0.01G
~ 0.1G
(3% + 3 digits)
(3% + 3 digits) reading;
0.4%/G
1000V 0.00M~0.99G
1.00G~200G
0.01M
~ 0.01G
0.01G
~ 0.1G
(3% + 3 digits)
(3% + 3 digits) reading;
0.2%/G
Short-circuit current < 2mA
250V DC and AC rms input limit
Insulation Resistance Test Voltage Outputs
Output Voltage
Test Steps (50~120%)
Load
Overload
Test
current
Short-circuit
current
50V (0%~+20%)
(25~60V)
50k
Ω
250VAC/DC 1mA
≤
2mA
100V (0%~+20%)
(50~120V)
100k
Ω
250VAC/DC 1mA
≤
2mA
250V (0%~+20%)
(125~300V)
250k
Ω
250VAC/DC 1mA
≤
2mA