 Pocket InfraRed Thermometer 
 Model IR201A 

 

Introduction 

Congratulations on your purchase of Extech’s IR201A Pocket InfraRed 
Thermometer. This device is used to take non-contact temperature 
measurements simply by pointing the meter and pressing a button. This 
meter, with proper care, will provide years of safe reliable service. 

Description 

 

1 Infrared 

sensor 

2 Laser 

Pointer 

3 Measurement 

button 

4 Laser 

button 

5 Max/Set 

button 

Battery compartment (rear) 

7 LCD 

display 

Lanyard attachment point 

 

Setup 

The Setup mode allows for setting: Max or Min display, °C or °F display, 
Emissivity value, Lock continuous mode, High alarm limit and Alarm On/Off. 
Enter the Setup mode by taking a measurement then press the MAX button 
to step through and adjust the features. Each press of the MAX button 
advances to the next feature. 

Press the Measure button to turn the meter on. 

Press the MAX button to step to the Max/Min selection. Press the 
LASER button to set either MIN or MAX (blinking) and then press 
the Measure button to confirm the selection 

Press the MAX button to step to the C/F selection. Press the LASER 
button to set either C or F (blinking) and then press the Measure 
button to confirm the selection. 

Press the MAX button to step to the Ems (emissivity) increase 

 

selection. Press the LASER button to increase the 

ε

 (blinking) value 

and then press the Measure button to confirm the selection. 

Press the MAX button to step to the Ems (emissivity) decrease 

 

selection. Press the LASER button to decrease the 

ε

 (blinking) value 

and then press the Measure button to confirm the selection. 

Press the MAX button to step to the LOCK selection. Press the 
LASER button to set LOCK On and then Press & Hold the Measure 
button for two seconds to begin continuous measurements. Press 
the LASER button to cancel the LOCK mode. 

Press the MAX button to step to the Hl (High Limit) increase 

 

selection. Press the LASER button to increase the value and then 
press the Measure button to confirm the selection. 

Press the MAX button to step to the Hl (High Limit) decrease 

 

selection. Press the LASER button to decrease the value and then 
press the Measure button to confirm the selection. 

Press the MAX button to step to the Alarm On/Off selection. Press 
the LASER button to set either ON or OFF and then press the 
Measure button to confirm the selection. 

 
 

IR Measurements 

1. 

Point the meter toward the surface to be measured 

2. 

Press and hold the MEASURE button. The meter will turn ON, 
“SCAN” will blink and the measured temperature will be displayed. 

3. 

Release the MEASURE button and the last reading will be held 
(HOLD will appear) in the display for approximately 6 second and 
then the meter will turn off.  

4. 

In the measure mode, press the LASER button to turn the laser 
pointer on or off. 

5. 

Press the MAX button to display and hold the highest reading. The 
display will update with each higher reading. 

Measurement considerations:  

 

The thermometer must be acclimated to ambient temperature for 
accurate measurements. Allow approximately 30 minutes 
acclimation time when the storage area and use area have a large 
temperature differential. 

 

If known, adjust the emissivity value to match the emissivity of the 
surface being measured. If unknown, use 0.95. 

 

The surface measured should be cleaned of frost, dust or other 
materials before accurate measurements can be made.

  

 

Recommended measuring distance is 2” to 36” (5cm to 91cm)

 

 

Apply black tape or paint to highly reflective surfaces to increase the 
emissivity and improve the accuracy of the measurement. 

 

The object being measured should be larger than the spot size as 
illustrated in the distance to spot diagram. 

 

 

 

 

 

Battery Replacement 

The   battery icon appears when the batteries need replacing. Open the 
battery compartment on the rear of the meter, replace the batteries and 
close the compartment. 

 

You, as the end user, are legally bound (

EU Battery 

ordinance

) to return all used batteries, 

disposal in the 

household garbage is prohibited! 

You can dispose of your 

used batteries at collection points in your community or 
wherever batteries are sold! 

Disposal: 

Follow the valid legal stipulations in respect of the 

disposal of the device at the end of its lifecycle 

 

Specifications 

Display 

3 ½ -digit (2000 count) LCD display 

Power off 

Automatic shutoff after approx. 8 seconds 

Laser power 

Laser power less than 1mW (red) 

Distance to Target Ratio 

6:1 

Emissivity settings 

0.10 to 1.0 adjustable 

Operating Temperature 

0 to 50°C (32 to 122°F) 

Operating Humidity 

Max. 80% RH 

Power Supply 

2 x AAA battery 

Weight 

76g / 2.7 oz. 

Size 

94 x 51 x 25 mm (3.7 x 2 x 1") 

 

Range Specifications

 

Ranges(Resolution)  -30 to 270°C / -22 to 518

o

F (0.1°C/F) 

Accuracy 

 ± (2.5% of r 2ºC or 4ºF) 

 

 
 