8 GX900-EU-EN V1.6 4/14
SAFETY
INSTRUCTIONS
–
INPUT
PROTECTION
LIMITS
This
meter
has
been
designed
for
safe
use,
but
must
be
operated
with
caution.
The
rules
listed
in
this
safety
section
must
be
carefully
followed
for
safe
operation.
1.
NEVER
apply
voltage
or
current
to
the
meter
that
exceeds
the
specified
maximum:
Input
Protection
Limits
Function
Maximum
Input
V
DC
or
V
AC
1000V
DC/AC
rms
mA
AC/DC
800mA
1000V
fast
acting
fuse
A
AC/DC
10A
1000V
fast
acting
fuse
Frequency,
Resistance,
Capacitance,
Duty
Cycle,
Diode
Test,
Continuity
1000VDC/AC
rms
Temperature
1000VDC/AC
rms
2.
USE
EXTREME
CAUTION
when
working
with
high
voltages.
3.
DO
NOT
measure
voltage
if
the
voltage
on
the
"COM"
input
jack
exceeds
1000V
above
earth
ground.
4.
NEVER
connect
the
meter
leads
across
a
voltage
source
while
the
function
switch
is
in
the
current,
resistance,
or
diode
mode.
Doing
so
can
damage
the
meter.
5.
ALWAYS
discharge
filter
capacitors
in
power
supplies
and
disconnect
the
power
when
making
resistance
or
diode
tests.
6.
ALWAYS
turn
off
the
power
and
disconnect
the
test
leads
before
opening
the
covers
to
replace
the
fuse
or
batteries.
7.
NEVER
operate
the
meter
unless
the
back
cover
and
the
battery
and
fuse
covers
are
in
place
and
fastened
securely.
8.
If
the
equipment
is
used
in
a
manner
not
specified
by
the
manufacturer,
the
protection
provided
by
the
equipment
may
be
impaired.