Measurement and Display Considerations

 

If the unit appears to be locked (display frozen). It is possible that the Data Hold mode has 
been inadvertently accessed by pressing the 

MODE/HOLD

 button. (“HOLD” will be displayed 

in the bottom left of the LCD.) Simply press the MODE button again or turn the meter off and 
then on. 

 

For maximum accuracy, allow sufficient time for the temperature of the probe to reach the 
temperature of the sample before calibrating. This will be indicated by a stable temperature 
reading on the display. 

Reset Calibration Data 

Follow this procedure to clear all calibration data from the meter.  Resetting the calibration data may 
be necessary when new calibration solutions are used or accuracy of measurements is in question. 

1. Turn off the meter. 
2. Press and Hold the Cal/Recall and Mode/Hold buttons. 
3. Momentarily press the On/Off button, as soon as the display comes on, release all 3 

buttons.

4. The display will show  “

dFLt rSt

” (default reset) and all of the calibration data will be erased. 

If  “

dFLt  rSt”

 does not appear, retry the procedure. 

5. Proceed to the calibration routine for pH and Conductivity. 

Calibration - Conductivity 

Meter accuracy verification should be performed on a periodic basis. Once per month is the 
recommended cycle for normal use. If calibration is required, a conductivity standardizing solution 
must be obtained. The meter can be calibrated in any or all of the three ranges. Standardizing 
solutions of 84µS/cm, 1413µS/cm or 12.88mS/cm (12,880µS/cm) are used for the automatic 
calibration recognition procedure. No other calibration values are permitted. 
Calibration is always done in conductivity mode. Since salinity and TDS values are calculated from 
conductivity values, this procedure also calibrates the salinity and TDS ranges. 

1. 

Fill a sample cup with the standardizing solution. 

2. 

Turn the meter ON and insert the electrode into the solution. Tap or move the electrode in the 
sample to dislodge any air bubbles. 

3. 

Press and hold the 

CAL/RECALL

 button (approximately 2 seconds) until “CAL” appears in 

the lower (temp) display. The main display will start flashing. 

4. 

The meter will automatically recognize and calibrate to the standardizing solution. The display 
will briefly indicate “SA”, “End” and then return to the measurement mode after a calibration. 
Note:  The “SA” will not appear if the calibration fails. 

5. 

The “range calibrated” symbol will appear in the display for each range that is calibrated 
during that power on cycle. 

Low range, 84µS/cm 
Medium range, 1413µS/cm 
High range, 12.88mS/cm (12,880µS/cm) 

Note: Each time the calibration mode is entered all calibration symbols on the display are cleared, 

but only the calibration data for the currently calibrated range is replaced. The other two ranges 
keep the existing calibration data, just the symbols are removed. Calibration of all three ranges must 
be performed during one power on period for all three range calibration symbols to appear.  

See

Reset Calibration Data

 to clear all calibration data from the meter. 

Note: The meter allows for a 1, 2 or 3 point calibration. If calibration is done for more than one point 
the lowest value standard should be done first to obtain the best accuracy. 

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for ExStik EC400

Page 1: ...User Manual ExStik II EC400 Conductivity TDS Salinity Temperature Meter www com information itm com 1 800 561 8187 ...

Page 2: ...f the batteries are weak the BAT indicator appears on the LCD Press the ON OFF key to turn the ExStik II on or off The auto power off feature shuts the ExStik II off automatically after 10 minutes of inactivity to preserve battery life Getting Started Remove the cap from the bottom of the ExStik II to expose the conductivity electrode Before the first use or after storage rinse the electrode in de...

Page 3: ...e mode Hold and store data 4 CAL RECALL button calibrate change temperature units recall data 5 ON OFF button 6 Electrode Collar 7 Electrode LCD Display 1 Bargraph display 2 Measurement units 3 Main display 4 Range calibration and low battery indicators 5 Temperature display 6 Reading Hold Indicator www com information itm com 1 800 561 8187 ...

Page 4: ...ess the ON button and then SELF CAL will appear in the display during the turn on diagnostics 2 Depress and hold the MODE HOLD key to scroll to the desired measurement mode 3 Insert the electrode into the sample making sure that the electrodes are completely submersed 4 Slowly stir the solution with the electrode to remove air bubbles 5 The meter will auto range to the proper range and then displa...

Page 5: ...isplay when the meter is first turned on or when changing measurement function to TDS Note In the Salinity mode the ratio is 0 4 to 0 6 automatic To change the ratio while in the TDS measurement mode ppm or mg l 1 Press and release the CAL RECALL button twice in succession The stored ratio will appear in the display 2 Press the MODE HOLD button to increase the ratio value in steps of 0 1 3 When th...

Page 6: ...ECALL button depressed momentarily press the ON OFF button When SELF CAL appears in the display release the CAL RECALL button The unit will power on with temperature displayed in the new units Data Hold Mode Press the MODE HOLD button to hold freeze a reading in the display The meter will enter the HOLD mode and the HOLD indicator will appear Note This also stores the reading Press the MODE HOLD b...

Page 7: ...or the automatic calibration recognition procedure No other calibration values are permitted Calibration is always done in conductivity mode Since salinity and TDS values are calculated from conductivity values this procedure also calibrates the salinity and TDS ranges 1 Fill a sample cup with the standardizing solution 2 Turn the meter ON and insert the electrode into the solution Tap or move the...

Page 8: ... Momentary press of the MODE HOLD key Displays last in first out Exit Memory Retrieval On Memory Recall Momentary press of the CAL RECALL key Clear Stored Memory On Any Measure Mode Press and hold the ON OFF key for 4 seconds until clr is displayed Change Measurement Mode On Any Press and hold the MODE HOLD key for at least 2 seconds the modes will scroll by until the key is released Enter Cond TD...

Page 9: ...0 0o C 32 0 F to 140o F Temperature Range 0 0 C to 65 0o C 32 0 F to 149 F Temperature Resolution 0 1 up to 99 9 1 100 Temperature Accuracy 1 C 1 8 F from 0 to 50 C 32 to 122 F 3 C 5 4 F from 50 to 65 C 122 to 149 F Accuracy Conductivity 2 full scale TDS 2 full scale Salinity 2 full scale Measurement Memory 25 tagged numbered readings Low battery indication BAT appears on the LCD Power Four 4 CR20...

Page 10: ...ld Disposal Do not dispose of this instrument in household waste The user is obligated to take end of life devices to a designated collection point for the disposal of electrical and electronic equipment Electrode Replacement 1 To remove an electrode unscrew and completely remove the electrode collar turn the collar counter clockwise to remove 2 Gently rock the electrode from side to side pulling ...

Page 11: ... MODE HOLD key to exit HOLD mode BAT message Batteries are low Replace batteries Unit will not calibrate in conductivity mode Contaminated conductivity standards Use fresh standards Unit will not calibrate in conductivity mode Dirty probe Clean conductivity probe See cleaning instructions Unit will not calibrate in conductivity mode Damaged conductivity probe Replace electrode Unit will not calibr...

Reviews: