EX320-EU-ENG V5.1 6/09
7
AC / DC CURRENT MEASUREMENTS
CAUTION:
Do not make current measurements at 10 Amps for longer than 30
seconds. Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1.
Insert the black test lead banana plug into the negative
COM
jack.
2.
For current measurements up to 4000µA, set the function switch to the µA position
and insert the red test lead banana plug into the
mA/µA
jack
3.
For current measurements up to 400mA, set the function switch to the mA position
and insert the red test lead banana plug into the
mA/µA
jack.
4.
For current measurements up to 10A, set the function switch to the 10A range and
insert the red test lead banana plug into the
10A
jack.
5.
Use the MODE button to select AC or DC current. The display will reflect the selection.
6.
Remove power from the circuit under test, then open up the circuit at the point where
you wish to measure current.
7.
Touch the black test probe tip to the negative side of the circuit. Touch the red test
probe tip to the positive side of the circuit.
8.
Apply power to the circuit.
9.
Read the current in the display.