Air Flow (CFM, CMM) Volume Measurements 

 

Press and hold the 

M

 button until ‘AirFl’ appears. 

 

Enter the duct area (in square feet or meters) by short pressing 

M

 and long pressing 

R

 until 

‘m-2’ (m

2

) or ‘f-2’ (ft

2

) appears. Short press the power button to increase or 

H

 to decrease 

the area; Long press the power or 

H

 button to scroll quickly. See area equations below. 

 

Short press   to save. 

 

Remember to convert inches, centimeters, or other unit to square feet or square meters 

before entering the area; otherwise, the air flow readings will not be correct. 

 

Place the vane in the air duct and read the air flow reading in CFM (ft

3

) or CMM (m

3

) units on 

the LCD. 

 

Area equation for rectangular or square ducts 

 
 

                              

 

 
Area equation for circular ducts 

 
 
 

 
 

Cubic equations 

CFM (ft

3

/min) = Air Velocity (ft/min) x Area (ft

2

CMM (m

3

/min) = Air Velocity (m/s) x Area (m

2

) x 60 

 

 

 

 
 

 

 

 

 

Width (W) 

Height (H) 

Area (A) = Width (W) x Height (H) 

Radius (r) 

Area (A)

 = 

π

 

x

 

r

2

 

Where

 

π

 

3.14

 

and r

2

 = 

radius x radius

 

Summary of Contents for EN510

Page 1: ...USER MANUAL ENVIRONMENTAL METER Model EN510 Additional User Manual Translations available at www extech com ...

Page 2: ...with air temperature readings Air Flow Volume measurements in CFM ft3 and CMM m3 units Light measurements in Foot candle and LUX units Environmental measurements Relative humidity with air temperature Dew point temperature Wet bulb temperature Wind chill temperature Heat Index temperature and Type K temperature with external probe connected Low friction ball bearing mounted vane wheel for high acc...

Page 3: ...Temperature sensors 7 Power Backlight button 8 MAX MIN Record Enter Button Lanyard not pictured Battery compartment and tripod mount on rear of meter LCD Description 1 MAX MIN Record icon 2 MAX Display 3 MIN Display 4 Display Hold 5 Primary Display 6 Units of Measure 7 Temperature Display 8 Reverse Orientation Display 9 Units of Measure for reverse display ...

Page 4: ...ay icons with brief descriptions below An Anemometer Air Velocity plus Air temperature AirFL Air flow Volume in CFM and CMM units CHiLL Wind chill temperature rH Relative Humidity in and Air temperature dP Dew point temperature _Et Wet bulb temperature HEAt Heat index temperature LigHt Light intensity measurements tYPE Type K thermocouple temperature must attach thermocouple to meter Changing the ...

Page 5: ... to toggle MAX and MIN reading displays Short press the H button to reset MAX and MIN memories Press and hold the R button to exit the Recording mode Measurement Modes Air Velocity with Air temperature With the meter ON press and hold the M Mode button until An is displayed then release the button Hold the measurement vane in the air stream in either direction View the air velocity and air tempera...

Page 6: ...low Short press to save Remember to convert inches centimeters or other unit to square feet or square meters before entering the area otherwise the air flow readings will not be correct Place the vane in the air duct and read the air flow reading in CFM ft3 or CMM m3 units on the LCD Area equation for rectangular or square ducts Area equation for circular ducts Cubic equations CFM ft3 min Air Velo...

Page 7: ...he meter is not to be used for 60 days or more remove the batteries and store separately Never dispose of used batteries or rechargeable batteries in household waste As consumers users are legally required to take used batteries to appropriate collection sites the retail store where the batteries were purchased or wherever batteries are sold Disposal Do not dispose of this instrument in household ...

Page 8: ...ure requires external thermocouple Tripod mount On rear of meter Operating Humidity 80 RH Max Operating Temperature 0 to 50 C 32 to 122 F Over Limit Display Power Supply 3 x 1 5V AAA batteries Power Consumption Approximately 5mA DC Weight 160g 5 6oz Dimensions HxWxD 153x 58x 25mm 6 0 x 2 3 x 1 0 inch Air Velocity plus Air Temperature Units Range Resolution Accuracy ft min FPM 80 to 3937 1 3 F S m ...

Page 9: ...ature Unit Range Resolution Accuracy C 25 3 to 49 0 0 1 Calculated from temperature and humidity readings F 13 5 to 120 0 0 1 Wet Bulb Temperature C 5 4 to 49 0 0 1 Calculated from temperature and humidity readings F 22 2 to 120 0 1 Heat Index C 0 to 100 0 0 1 2 0 C F 32 to 212 0 1 3 6 F Exposure to direct sunlight can increase heat index readings by 8o C 14o F Type K Thermometer C 50 0 to 1300 0 ...

Page 10: ...it of Measure Conversion Table m s ft min knots km h MPH 1 m s 1 196 87 1 944 3 6 2 24 1 ft min 0 00508 1 0 00987 0 01829 0 01138 1 knot 0 5144 101 27 1 1 8519 1 1523 1 km h 0 2778 54 69 0 54 1 0 6222 1 MPH 0 4464 87 89 0 8679 1 6071 1 Copyright 2017 FLIR Systems Inc All rights reserved including the right of reproduction in whole or in part in any form ISO 9001 Certified www extech com ...

Reviews: